Multilumen central venous catheters in children: relative potential to perforate vessels. An in vitro study.

R H Welch1, N Gravenstein, R H Blackshear.   

Abstract

OBJECTIVE: Because cardiovascular perforation by a central venous catheter (CVC) is a serious complication of catheterization in pediatric patients, we conducted an in vitro study of the relative potential for perforation of a standard material by the tips of multilumen pediatric catheters. Since we could not simulate vessel tissue, we hypothesized that testing catheters on a standard material would show whether catheters varied in tendency to perforate such a material and thus indicate a "relative potential for perforation."
METHODS: Each CVC protruding from a support tube was suspended in a water-filled Plexiglas chamber at a 90 degrees incident angle to a polyethylene film, which was made to bulge 6 mm into the CVC tip 120 times per minute by hydropressure. Perforation of the polyetheylene film was documented on a time-based, strip-chart recording of pressure change on the opposite side of the film. We recorded the number of pulsations required for the following catheters to perforate the polyethylene: Arrow flex tip, Cook polyurethane, Viggo hydrocath polyurethane, and Cook silicone CVCs of 4- and 5-Fr size with 2 or 3 lumens (n = 5 catheters of each type, each catheter being tested 5 times).
RESULTS: The number of pulsations to perforation ranged from 1 +/- 0.4 SD to > 7000.
CONCLUSIONS: This in vitro study of the worst-case condition (90 degrees incident angle between CVC tip and polyethylene film) indicates that pediatric multilumen CVCs vary significantly in their relative potential to perforate a standard material. We suggest that, when central venous catheterization is contemplated in children, in addition to insertion site, catheter length, and depth of insertion, the type of catheter is another variable to consider in order to minimize the chance of cardiovascular perforation by the CVC tip.

Entities:  

Mesh:

Year:  1997        PMID: 9112202     DOI: 10.1023/a:1007389228641

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  20 in total

Review 1.  Complications of central venous catheterization.

Authors:  S E Mitchell; R A Clark
Journal:  AJR Am J Roentgenol       Date:  1979-09       Impact factor: 3.959

2.  Perforation of the heart or vena cava by central venous catheters inserted for monitoring or infusion therapy.

Authors:  A W Jay; H E Aldridge
Journal:  CMAJ       Date:  1986-11-15       Impact factor: 8.262

3.  The consequences of a cavalier approach to central venous catheterization.

Authors:  V L Hoshal
Journal:  Acta Anaesthesiol Scand Suppl       Date:  1985

4.  A short history of central venous catheterization.

Authors:  E Kalso
Journal:  Acta Anaesthesiol Scand Suppl       Date:  1985

Review 5.  Cardiac tamponade from central venous catheters.

Authors:  H Edwards; T C King
Journal:  Arch Surg       Date:  1982-07

6.  Migration of tips of central venous catheters in seated patients.

Authors:  D S Lee; J Kuhn; M J Shaffer; H D Weintraub
Journal:  Anesth Analg       Date:  1984-10       Impact factor: 5.108

7.  Percutaneous catheterization of the internal jugular vein in infants and children.

Authors:  S R Prince; R L Sullivan; A Hackel
Journal:  Anesthesiology       Date:  1976-02       Impact factor: 7.892

8.  Total intravenous hyperalimentation (TIH) complications in childhood: a radiological survey.

Authors:  F Bellini; G Beluffi; N Principi
Journal:  Pediatr Radiol       Date:  1984

9.  Catheter-induced lesions of the right side of the heart. A one-year prospective study of 141 autopsies.

Authors:  B S Ducatman; J C McMichan; W D Edwards
Journal:  JAMA       Date:  1985-02-08       Impact factor: 56.272

10.  Percutaneous central venous catheterization in a pediatric intensive care unit: a survival analysis of complications.

Authors:  J P Stenzel; T P Green; B P Fuhrman; P E Carlson; R P Marchessault
Journal:  Crit Care Med       Date:  1989-10       Impact factor: 7.598

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.