| Literature DB >> 9098975 |
S R Bakalyar1, C Phipps, B Spruce, K Olsen.
Abstract
The relationship between sample volume and the chromatographic performance factors of detection sensitivity and resolution were explored for high-performance liquid chromatography (HPLC) columns of 1.0, 2.1 and 4.6 mm I.D. Performance was evaluated for isocratic and gradient elution in systems having low, intermediate and high extracolumn dispersion. Sample volumes ranged from 0.4 microliter to 2 ml. This paper provides guideline for selecting sample injection parameters so that the performance of microbore columns is not needlessly sacrificed by using too small or too large a sample volume. The guidelines are extracted from the works of others and are modified by our own experimental results. Comments are also made on the relative merits of various injection techniques.Mesh:
Year: 1997 PMID: 9098975 DOI: 10.1016/s0021-9673(96)00851-5
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759