Literature DB >> 9095699

An assessment of the effectiveness of the revised FDA checklist.

R Manley1, J D Cuddeford.   

Abstract

In 1978, the Food and Drug Administration (FDA), developed a generic anesthesia equipment preuse checklist. The checklist was first released by the FDA in August 1986 and endorsed by the American Association of Nurse Anesthetists on October 18, 1986. The FDA checklist was revised in 1992 to improve the abilities of anesthesia providers to detect machine faults. In the present study, the investigators attempted to determine the effectiveness of the revised FDA checklist in detection of anesthesia machine faults as compared to providers' usual methods. Whereas no published study of preanesthesia safety inspection had been performed since the revision of the FDA checklist, the authors compared the detection abilities of anesthesia providers before and after inclusion of the revised FDA checklist. Twenty-two anesthesia providers were tested to compare the number of prearranged anesthesia machine faults that could be detected with (1) their usual checkout methods, and (2) with the revised FDA checklist. Data describing the subjects' fault detection abilities were analyzed using the t test for paired observation (P value < 0.05 considered significant). Statistical analysis revealed no significant difference (P = 0.479) when subjects used the FDA checklist and when they used their usual method. Use of the FDA machine checklist was no more effective than the provider's usual method in discovering machine faults. When using their normal method, 54.5% of providers did not discover more than 50% of programmed faults. Approximately 40.9% of providers who used the revised FDA checklist did not discover over 50% of programmed faults.

Mesh:

Year:  1996        PMID: 9095699

Source DB:  PubMed          Journal:  AANA J        ISSN: 0094-6354


  4 in total

1.  Can a structured checklist prevent problems with laparoscopic equipment?

Authors:  E G G Verdaasdonk; L P S Stassen; W F Hoffmann; M van der Elst; J Dankelman
Journal:  Surg Endosc       Date:  2008-07-03       Impact factor: 4.584

2.  Impact of repeated simulation on learning curve characteristics of residents exposed to rare life threatening situations.

Authors:  Sree Kumar E J; Makani Purva; Sarat Chander M; Aruna Parameswari
Journal:  BMJ Simul Technol Enhanc Learn       Date:  2020-11-01

Review 3.  Requirements for the design and implementation of checklists for surgical processes.

Authors:  E G G Verdaasdonk; L P S Stassen; P P Widhiasmara; J Dankelman
Journal:  Surg Endosc       Date:  2008-07-18       Impact factor: 4.584

4.  Pre-use anesthesia machine check; certified anesthesia technician based quality improvement audit.

Authors:  Mazen Al Suhaibani; Assaf Al Malki; Saad Al Dosary; Hanan Al Barmawi; Mahdhav Pogoku
Journal:  Anesth Essays Res       Date:  2014 Sep-Dec
  4 in total