Literature DB >> 9095611

[Breakage of a Seldinger spring guide wire during percutaneous catheterization of a subclavian vein].

T Shimamoto1, T Arai.   

Abstract

Spring guide wires have been widely used for the central venous catheterization with the popularity of the Seldinger technique. We report here an accidental breakage of a spring guide wire during percutaneous catheterization of a subclavian vein. The venipuncture by a plastic catheter over an introducer needle and the insertion of a spring guide wire through the catheter were easily achieved. However, the insertion of an indwelling catheter over the guide wire was difficult to perform, and the pulling out of the guide wire from the catheter was more difficult, which brought the uncoiling of the spring part of the guide wire from the mandrel. The tip of the guide wire was not cut off, and there were no sequelae in the patient. A plastic catheter over an introducer needle for a subclavian venipuncture is so flexible that it is occasionally bent between the clavicle and the first rib after pulling out of the inside needle. Probably, that is the reason of the difficulties of the insertion of an indwelling catheter and the pulling out of a guide wire. The use of a rigid metal needle for the venipuncture is an alternative way to avoid these troubles.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9095611

Source DB:  PubMed          Journal:  Masui        ISSN: 0021-4892


  4 in total

1.  Complications of Central Venous Catheterisation: Breakage of guidewire-a disaster averted.

Authors:  Qutaiba A Tawfic; Pradipta Bhakta; Jyoti Burad; Pragyandipta Mishra; Rajini Kausalya
Journal:  Sultan Qaboos Univ Med J       Date:  2011-10-25

2.  Development of a safe guidewire.

Authors:  Toshiyasu Suzuki; Kenji Ito; Junichi Nishiyama; Keiichiro Hasegawa; Masahiro Kanazawa; Haruo Fukuyama
Journal:  J Anesth       Date:  2006       Impact factor: 2.078

3.  Fracture of J-tipped guidewire during central venous catheterization and its successful removal under fluoroscopic guidance -A case report-.

Authors:  Sun Kyung Park; In-Kyong Yi; Jae-Ho Lee; Dae-Hee Kim; Sook Young Lee
Journal:  Korean J Anesthesiol       Date:  2012-11-16

4.  Jugular Venous Catheterization: A Case of Knotting.

Authors:  E Erkılıç; M Aksoy; T Gümüş; S Sarı; E Kesimci
Journal:  Case Rep Med       Date:  2015-11-08
  4 in total