Technical factors of CT angiography studied with a carotid artery phantom.

S W Wise1, K D Hopper, T A Schwartz, T R Ten Have, C J Kasales.   

Abstract

PURPOSE: To evaluate scanning parameters (conventional versus spiral CT, section thickness, and pitch) and vessel orientation in the performance of CT angiography.
METHODS: Conventional CT and 1.0-, 1.5-, and 2.0-pitch spiral CT acquisitions of a carotid phantom designed with vessels oriented parallel to the z-axis, 45 degrees oblique, and perpendicular to the z-axis were obtained with section thicknesses of 2, 4, and 8 mm. The phantom contained 32 vessels with 0% to 100% stenoses. Normal and stenotic luminal diameters were measured and the number of artifacts was assessed.
RESULTS: No overall difference was observed among conventional and spiral CT acquisitions obtained with pitches of 1.0, 1.5, and 2.0. With thicker sections, CT angiographic accuracy decreased and artifacts increased. The three-vessel orientations were relatively comparable in accuracy in terms of the percentage of stenosis measured. Vessels parallel to the z-axis suffered less artifactual degradation. Unique artifacts, such as luminal distortion and beam hardening, were observed in vessels oriented at 45 degrees and perpendicular to the z-axis.
CONCLUSION: Use of thinner sections with vessels oriented parallel to the z-axis optimizes CT angiographic quality. There is no apparent degradation with the use of spiral CT, and a pitch of 1.5 or 2.0 provides results equivalent to 1.0-pitch spiral studies.

Mesh:

Year:  1997        PMID: 9090394      PMCID: PMC8338401     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  2 in total

Review 1.  Brain imaging.

Authors:  R I Grossman
Journal:  AJNR Am J Neuroradiol       Date:  2000-01       Impact factor: 3.825

2.  Changes in measured size of atherosclerotic plaque calcifications in dual-energy CT of ex vivo carotid endarterectomy specimens: effect of monochromatic keV image reconstructions.

Authors:  Lorenzo Mannelli; Lee M Mitsumori; Marina Ferguson; Dongxiang Xu; Baocheng Chu; Kelley R Branch; William P Shuman; Chun Yuan
Journal:  Eur Radiol       Date:  2012-08-21       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.