Management software for a universal device communication controller: application to monitoring and computerized infusions.

E J Coussaert1, F R Cantraine.   

Abstract

We designed a virtual device for a local area network observing, operating and connecting devices to a personal computer. To keep the widest field of application, we proceeded by using abstraction and specification rules of software engineering in the design and implementation of the hardware and software for the Infusion Monitor. We specially built a box of hardware to interface multiple medical instruments with different communication protocols to a PC via a single serial port. We called that box the Universal Device Communication Controller (UDCC). The use of the virtual device driver is illustrated by the Infusion Monitor implemented for the anaesthesia and intensive care workstation.

Mesh:

Substances:

Year:  1996        PMID: 9080243     DOI: 10.1007/bf02919399

Source DB:  PubMed          Journal:  Int J Clin Monit Comput        ISSN: 0167-9945


  1 in total

1.  The first object oriented monitor for intravenous anesthesia.

Authors:  F R Cantraine; E J Coussaert
Journal:  J Clin Monit Comput       Date:  2000-01       Impact factor: 2.502