Literature DB >> 9068596

Comparative evaluation of three assay systems for automated determination of hemoglobin A1c.

G Halwachs-Baumann1, S Katzensteiner, W Schnedl, P Pürstner, T Pieber, M Wilders-Truschnig.   

Abstract

We evaluated three newly introduced systems for automated determinations of hemoglobin (Hb) A1c, which allow the processing of large amounts of samples in a routine clinical laboratory. We compared these methods--the Variant HPLC, the Hi-Auto A1c analyzer system, and the Roche immunoassay--with the Diamat HPLC system. All showed good precision and good concordance with the Diamat HPLC. The reference range for Hb A1c has to be determined by the laboratory for each assay system. Interference study showed no statistically significant influence of anemia, polycythemia, rheumatoid factor, or chronic hemodialysis, although individual Hb A1c values can be influenced by polycythemia (when measured with the Hi-Auto A1c analyzer) and by chronic hemodialysis (when measured with the Variant HPLC). HPLC was not suitable for measuring Hb A1c in the examined cases of hemoglobin variants; assaying fructosamine seems to be better for monitoring these patients.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9068596

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

1.  Evaluation of the performance and stability of a novel A1c-cellular control.

Authors:  Kausik Das; Gary D Krzyzanowski; Stephanie M Wigginton; Joel M Lechner; Wayne L Ryan
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Hemoglobin A1c levels in non-diabetic patients with end-stage renal disease (ESRD) receiving hemodialysis.

Authors:  X Wang; S K Peesapati; M F Renedo; S Moktan
Journal:  J Endocrinol Invest       Date:  2004-09       Impact factor: 4.256

3.  Degradation products of proteins damaged by glycation, oxidation and nitration in clinical type 1 diabetes.

Authors:  N Ahmed; R Babaei-Jadidi; S K Howell; P J Beisswenger; P J Thornalley
Journal:  Diabetologia       Date:  2005-06-30       Impact factor: 10.122

4.  Crystallization and preliminary crystallographic analysis of two eukaryotic fructosyl peptide oxidases.

Authors:  Atsushi Ichiyanagi; Kozo Hirokawa; Keiko Gomi; Toru Nakatsu; Hiroaki Kato; Naoki Kajiyama
Journal:  Acta Crystallogr Sect F Struct Biol Cryst Commun       Date:  2013-01-30

5.  Long-term evaluation of a structured outpatient education programme for intensified insulin therapy in patients with Type 1 diabetes: a 12-year follow-up.

Authors:  J Plank; G Köhler; I Rakovac; B M Semlitsch; K Horvath; G Bock; B Kraly; T R Pieber
Journal:  Diabetologia       Date:  2004-07-17       Impact factor: 10.122

6.  Diabetes control improved when inner-city patients received graphic feedback about glycosylated hemoglobin levels.

Authors:  Rebecca B Chapin; Derrick C Williams; Richard F Adair
Journal:  J Gen Intern Med       Date:  2003-02       Impact factor: 5.128

7.  Hemoglobin determination with paired emitter detector diode.

Authors:  Elżbieta Mieczkowska; Robert Koncki; Łukasz Tymecki
Journal:  Anal Bioanal Chem       Date:  2010-11-03       Impact factor: 4.142

8.  Comparative evaluation of three different methods for HbA1c measurement with High-performance liquid chromatography in diabetic patients.

Authors:  Azadeh Karami; Azar Baradaran
Journal:  Adv Biomed Res       Date:  2014-03-25
  8 in total