Literature DB >> 9066326

Influence of thermoregulatory vasomotion and ambient temperature variation on the accuracy of core-temperature estimates by cutaneous liquid-crystal thermometers.

T Ikeda1, D I Sessler, D Marder, J Xiong.   

Abstract

BACKGROUND: Recently, liquid crystal skin-surface thermometers have become popular for intraoperative temperature monitoring. Three situations during which cutaneous liquid-crystal thermometry may poorly estimate core temperature were monitored: (1) anesthetic induction with consequent core-to-peripheral redistribution of body heat, (2) thermoregulatory vasomotion associated with sweating (precapillary dilation) and shivering (minimal capillary flow), and (3) ambient temperature variation over the clinical range from 18-26 degrees C.
METHODS: The core-to-forehead and core-to-neck temperature difference was measured using liquid-crystal thermometers having an approximately 2 degrees C offset. Differences exceeding 0.5 degree C (a 1 degree C) temperature range) were a priori deemed potentially clinically important. Seven volunteers participated in each protocol. First, core-to-peripheral redistribution of body heat was produced by inducing propofol/desflurane anesthesia; anesthesia was then maintained for 1 h with desflurane. Second, vasodilation was produced by warming unanesthetized volunteers sufficiently to produce sweating; intense vasoconstriction was similarly produced by cooling the volunteers sufficiently to produce shivering. Third, a canopy was positioned to enclose the head, neck, and upper chest of unanesthetized volunteers. Air within the canopy was randomly set to 18, 20, 22, 24, and 26 degrees C.
RESULTS: Redistribution of body heat accompanying induction of anesthesia had little effect on the core-to-forehead skin temperature difference. However, the core-to-neck skin temperature gradient decreased approximately 0.6 degree C in the hour after induction of anesthesia. Vasomotion associated with shivering and mild sweating altered the core-to-skin temperature difference only a few tenths of a degree centigrade. The absolute value of the core-to-forehead temperature difference exceeded 0.5 degree C during approximately 35% of the measurements, but the difference rarely exceeded 1 degree C. The core-to-neck temperature difference typically exceeded 0.5 degree C and frequently exceeded 1 degree C. Each 1 degree C increase in ambient temperature decreased the core-to-fore-head and core-to-neck skin temperature differences by less than 0.2 degree C.
CONCLUSIONS: Forehead skin temperatures were better than neck skin temperature at estimating core temperature. Core-to-neck temperature differences frequently exceeded 1 degree C (a 2 degrees C range), whereas two thirds of the core-to-forehead differences were within 0.5 degree C. The core-to-skin temperature differences were, however, only slightly altered by inducing anesthesia, vasomotor action, and typical intraoperative changes in ambient temperature.

Entities:  

Mesh:

Year:  1997        PMID: 9066326     DOI: 10.1097/00000542-199703000-00012

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  5 in total

1.  Control of body temperature and immune function in patients undergoing open surgery for gastric cancer.

Authors:  Li Shao; Nannan Pang; Ping Yan; Fengju Jia; Qi Sun; Wenjuan Ma; Yi Yang
Journal:  Bosn J Basic Med Sci       Date:  2018-08-01       Impact factor: 3.363

2.  Effects of season on sleep and skin temperature in the elderly.

Authors:  Kazue Okamoto-Mizuno; Kazuyo Tsuzuki
Journal:  Int J Biometeorol       Date:  2009-12-30       Impact factor: 3.787

3.  Unintended perioperative hypothermia.

Authors:  Stuart R Hart; Brianne Bordes; Jennifer Hart; Daniel Corsino; Donald Harmon
Journal:  Ochsner J       Date:  2011

Review 4.  Temperature monitoring and perioperative thermoregulation.

Authors:  Daniel I Sessler
Journal:  Anesthesiology       Date:  2008-08       Impact factor: 7.892

5.  Mass screening of suspected febrile patients with remote-sensing infrared thermography: alarm temperature and optimal distance.

Authors:  Ming-Fu Chiang; Po-Wei Lin; Li-Fong Lin; Hung-Yi Chiou; Ching-Wen Chien; Shu-Fen Chu; Wen-Ta Chiu
Journal:  J Formos Med Assoc       Date:  2008-12       Impact factor: 3.282

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.