In vitro bacterial adherence to hydrogel and poly(methyl methacrylate) intraocular lenses.

E W Ng1, G D Barrett, R Bowman.   

Abstract

PURPOSE: To compare the in vitro adherence of Staphylococcus epidermidis to poly-(hydroxyethyl methacrylate) (polyHEMA) or hydrogel intraocular lenses (IOLs) and poly(methyl methacrylate) (PMMA) IOLs.
SETTING: Lions Eye Institute, Perth, Western Australia.
METHODS: One-piece hydrogel lenses and one-piece PMMA lenses were suspended for 60 minutes in standardized suspensions of a well-characterized strain of S. epidermidis and then sonicated in a known quantity of balanced salt solution to remove the adherent bacteria. Quantitative cultures of the sonicates were performed and the results analyzed statistically.
RESULTS: The mean bacterial adherence of S. epidermidis to the PMMA IOLs (58,400 CFU) was more than 20 times greater than that to the hydrogel IOLs (1953 CFU). The difference was statistically significant (P < .001).
CONCLUSIONS: Adherence of S. epidermidis to hydrogel IOLs is significantly lower than to PMMA IOLs. This suggests that the risk of postoperative endophthalmitis after cataract extraction and IOL implantation may be lower with the use of hydrogel IOLs.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 9051525     DOI: 10.1016/s0886-3350(96)80093-9

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Effects of bacterial adhesion with respect to the type of material, structure and design of intraocular lenses.

Authors:  J I Alava; N Garagorri; N Briz; J Mendicute
Journal:  J Mater Sci Mater Med       Date:  2005-04       Impact factor: 3.896

2.  Risk factors of acute endophthalmitis after cataract extraction: a case-control study in Asian eyes.

Authors:  T Y Wong; S-P Chee
Journal:  Br J Ophthalmol       Date:  2004-01       Impact factor: 4.638

3.  Management and Microbiological Characteristics of Membrane Formation on a Hydrophilic Acrylic Intraocular Lens: A Clinical Case Series and Material Comparative Study of Different IOLs.

Authors:  Xiaodi Qiu; Yang Wu; Yongxiang Jiang; Yinghong Ji; Xiangjia Zhu; Jin Yang; Yi Lu
Journal:  J Ophthalmol       Date:  2019-02-03       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.