Literature DB >> 9050741

Genital tract burns during rollerball endometrial coagulation.

G A Vilos1, I D'Souza, D Huband.   

Abstract

Thermal injuries occurred in three women after hysteroscopic rollerball endometrial ablation. Possible mechanisms of injury included insulation failure of the rollerball to the sheath of the resectoscope, current division with the sheath acting as an alternate return electrode, and capacitative leakage of current from the active electrode to the sheath.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9050741     DOI: 10.1016/s1074-3804(97)80023-0

Source DB:  PubMed          Journal:  J Am Assoc Gynecol Laparosc        ISSN: 1074-3804


  1 in total

1.  Complications associated with monopolar resectoscopic surgery.

Authors:  George A Vilos; H Alshankiti; A G Vilos; B Asim Abu-Rafea; A Ternamian
Journal:  Facts Views Vis Obgyn       Date:  2020-05-07
  1 in total