Optimal conditions of 99mTc eluate for the radiolabeling of 99mTc-sestamibi.

J C Hung1, T J Herold, R J Gibbons.   

Abstract

Our nuclear pharmacy has reported that a failed radiochemical purity (RCP) (i.e., RCP < 90%) of 99mTc-sestamibi may be associated with the use of a first elution at later stages from a long-ingrowth time (i.e., > or = 72 h) wet-column generator. The primary purpose of this study was to evaluate the effects of 99mTc eluates from wet- and dry-column generators on the RCP of 99mTc-sestamibi under the above conditions. RCP values were found to be measurably higher and kit failure rates lower with the use of dry-column generator eluate. Using a dry-column generator eluate, Cardiolite kits were prepared with 11.10 GBq of 99mTc at 3, 4, and 5 h postelution and 5.55 GBq at 6, 10, 11, and 12 h postelution. Our data suggest that when 11.10 GBq of 99mTc from a dry-column generator with > or = 72-h ingrowth was used to prepare 99mTc-sestamibi, kit failure started to occur using 99mTc eluate at approximately 4 h postelution. When 5.55 GBq was used to reconstitute the kit, RCP failure began to occur using 99mTc eluate approximately 10 h postelution and wet-column generators; the failure rate can be reduced even further by avoiding the addition of high activities of 99mTc and long elution times.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 9044686     DOI: 10.1016/0969-8051(96)00054-6

Source DB:  PubMed          Journal:  Nucl Med Biol        ISSN: 0969-8051            Impact factor:   2.408


  1 in total

1.  Alternative chromatographic system for the quality control of lipophilic technetium-99m radiopharmaceuticals such as [(99m)Tc(MIBI)₆].

Authors:  D P Faria; C A Buchpiguel; F L N Marques
Journal:  Braz J Med Biol Res       Date:  2015-03-03       Impact factor: 2.590

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.