Evaluation of the IITC tail cuff blood pressure recorder in the rat against intraarterial pressure according to criteria for human devices.

M J Jamieson1, G M Gonzales, T I Jackson, S M Koerth, W F Romano, D X Tan, F Castillon, M H Skinner, M Grossmann, A M Shepherd.   

Abstract

National standards govern the manufacture and marketing of medical devices in the United States, including those for indirect blood pressure measurement in man. There are no comparable standards for devices for recording in laboratory animals. Noninvasive tail cuff blood pressure (BP) recording in the rat is widely accepted, but beset by methodologic difficulties. Intraarterial recording is regarded as the "gold standard" but is invasive and also susceptible to methodologic error. We compared the IITC Mark 12 photoelectric/oscillometric tail cuff system (IITC Life Sciences, Woodland Hills, CA) versus simultaneous femoral intraarterial recordings in spontaneously hypertensive rats, during anesthesia and 1 to 2 days after recover (150 recordings under each condition), according to the guidelines for human data collection and analysis suggested the American National Standard for automated sphygmomanometers. Within- and between-observer disagreements in estimates made by two observers from 40 anesthetized recordings were less for intraarterial measurements than for the tail cuff method. Within-observer differences (mean +/- SD of differences [SDD]) for systolic, diastolic, and mean pressure were 0 +/- 1, 0 +/- 1, and 0 +/- 1 mm Hg for intraarterial versus -1 +/- 3, 0 +/- 8, and 0 +/- 5 mm Hg for tail cuff. Between-observer differences were 0 +/- 2, 0 +/- 1, and 6 +/- 2 mm Hg versus 5 +/- 4, 13 +/- 7, and 0 +/- 5 mm Hg, respectively. Differences between tail cuff and intraarterial methods were 16 +/- 13, -5 +/- 11, and 2 +/- 8 mm Hg in anesthetized animals and 8 +/- 14, -5 +/- 9, and 0 +/- 9 mm Hg in conscious animals (39% to 82% of differences exceeded 5 mm Hg). The upper limits of clinically acceptable disagreement in the American National Standard are: mean of 5 mm Hg, SDD of 8 mm Hg. The disagreement between tail cuff and intraarterial recordings cannot be ascribed to either method with certainty. These findings do not support the manufacturer's guarantee of tail-cuff readings within "5 mm Hg of intraarterial." Inaccuracy and unreliability of devices intended for laboratory animal use have considerable scientific, fiscal, and ethical implications. Marketing of these devices should also be governed by rigorous standards.

Entities:  

Mesh:

Year:  1997        PMID: 9037330     DOI: 10.1016/s0895-7061(96)00321-4

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  8 in total

Review 1.  Developmental programming of the metabolic syndrome by maternal nutritional imbalance: how strong is the evidence from experimental models in mammals?

Authors:  James A Armitage; Imran Y Khan; Paul D Taylor; Peter W Nathanielsz; Lucilla Poston
Journal:  J Physiol       Date:  2004-09-30       Impact factor: 5.182

2.  Angiotensin AT2 receptor agonist prevents salt-sensitive hypertension in obese Zucker rats.

Authors:  Quaisar Ali; Sanket Patel; Tahir Hussain
Journal:  Am J Physiol Renal Physiol       Date:  2015-04-08

3.  Effects of salt loading on sympathetic activity and blood pressure in anesthetized two-kidney, one clip hypertensive rats.

Authors:  Besim Ozaykan; Ayşe Doğan
Journal:  Bosn J Basic Med Sci       Date:  2011-11       Impact factor: 3.363

4.  Angiotensin II Promotes Skeletal Muscle Angiogenesis Induced by Volume-Dependent Aerobic Exercise Training: Effects on miRNAs-27a/b and Oxidant-Antioxidant Balance.

Authors:  Luis Felipe Rodrigues; Bruno Rocha Avila Pelozin; Natan Daniel da Silva Junior; Ursula Paula Renó Soci; Everton Crivoi do Carmo; Glória de Fatima Alves da Mota; Victoria Cachofeiro; Vicente Lahera; Edilamar Menezes Oliveira; Tiago Fernandes
Journal:  Antioxidants (Basel)       Date:  2022-03-28

5.  Pitfalls of invasive blood pressure monitoring using the caudal ventral artery in rats.

Authors:  Hiroki Ohta; Takao Ohki; Yuji Kanaoka; Makoto Koizumi; Hirotaka J Okano
Journal:  Sci Rep       Date:  2017-02-13       Impact factor: 4.379

6.  Effects of Finerenone Combined with Empagliflozin in a Model of Hypertension-Induced End-Organ Damage.

Authors:  Peter Kolkhof; Elke Hartmann; Alexius Freyberger; Mira Pavkovic; Ilka Mathar; Peter Sandner; Karoline Droebner; Amer Joseph; Jörg Hüser; Frank Eitner
Journal:  Am J Nephrol       Date:  2021-06-10       Impact factor: 3.754

7.  Comparison of invasive blood pressure measurements from the caudal ventral artery and the femoral artery in male adult SD and Wistar rats.

Authors:  Ying Wang; Yushuang Cong; Jun Li; Xueting Li; Bing Li; Sihua Qi
Journal:  PLoS One       Date:  2013-04-05       Impact factor: 3.240

8.  Transmural Remodeling of Cardiac Microstructure in Aged Spontaneously Hypertensive Rats by Diffusion Tensor MRI.

Authors:  Archontis Giannakidis; Grant T Gullberg
Journal:  Front Physiol       Date:  2020-03-31       Impact factor: 4.566

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.