Determination of subnanogram concentrations of fentanyl in plasma by gas chromatography-mass spectrometry: comparison with standard radioimmunoassay.

B Fryirs1, A Woodhouse, J L Huang, M Dawson, L E Mather, B Fryirsa.   

Abstract

A method was devised to determine fentanyl plasma concentrations by GC-MS using selected-ion monitoring (SIM) with sufentanil as internal standard. This was compared with a commonly used commercial radioimmunoassay (RIA). Sample preparation for GC-MS involved basification of plasma then extraction using n-butyl chloride followed by concentration to dryness and reconstitution in toluene for chromatography. Using 1-ml plasma samples, the estimated limit of detection of fentanyl was 20 pg/ml. Blood samples for pharmacokinetic studies were split and assayed by GC-MS and RIA which had a limit of detection of 200 pg/ml. Pearson's r (r - 0.80, p < 0.0001) indicated the methods were highly correlated at all plasma concentrations. Owing to the greater sensitivity of the method, GC-MS is recommended over RIA for subnanogram determination of fentanyl in plasma.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9029316     DOI: 10.1016/s0378-4347(97)88058-9

Source DB:  PubMed          Journal:  J Chromatogr B Biomed Sci Appl        ISSN: 1387-2273


  2 in total

1.  Fatal outcome in a child after ingestion of a transdermal fentanyl patch.

Authors:  Jörg Teske; Jens-Peter Weller; Klaus Larsch; Hans Dieter Tröger; Matthias Karst
Journal:  Int J Legal Med       Date:  2006-11-18       Impact factor: 2.686

2.  Pulmonary administration of aerosolised fentanyl: pharmacokinetic analysis of systemic delivery.

Authors:  L E Mather; A Woodhouse; M E Ward; S J Farr; R A Rubsamen; L G Eltherington
Journal:  Br J Clin Pharmacol       Date:  1998-07       Impact factor: 4.335

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.