| Literature DB >> 9022175 |
P H Sharpe1, K Rajendran, J P Sephton.
Abstract
This paper describes work being carried out at the National Physical Laboratory towards the establishment of an alanine reference dosimetry service for radiotherapy applications. A precision fused quartz holder has been constructed to allow precise positioning of alanine dosimeters in the ESR cavity. A novel method of signal analysis based on spectrum fitting has been developed to minimize the effect of baseline distortions. Data are also presented on the relative response of alanine to 60Co gamma rays and high energy photons (4-12 MeV).Entities:
Mesh:
Substances:
Year: 1996 PMID: 9022175 DOI: 10.1016/s0969-8043(96)00174-1
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513