Literature DB >> 9018266

Meningitis after cochlear implantation in Mondini malformation.

E L Page1, T L Eby.   

Abstract

Although the potential for CSF leakage and subsequent meningitis after cochlear implantation in the malformed cochlea has been recognized, this complication has not been previously reported. We report a case of CSF otorhinorrhea and meningitis after minor head trauma developing 2 years after cochlear implantation in a child with Mondini malformation. Leakage of CSF was identified from the cochleostomy around the electrode of the implant, and this leak was sealed with a temporalis fascia and muscle plug. Although this complication appears to be rare, care must be taken to seal the cochleostomy in children with inner ear malformations at the initial surgery, and any episode of meningitis after surgery must be thoroughly investigated to rule out CSF leakage from the labyrinth.

Entities:  

Mesh:

Year:  1997        PMID: 9018266     DOI: 10.1016/S0194-59989770358-9

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   5.591


  2 in total

1.  Cochlear implant in incomplete partition type I.

Authors:  S Berrettini; F Forli; A De Vito; L Bruschini; N Quaranta
Journal:  Acta Otorhinolaryngol Ital       Date:  2013-02       Impact factor: 2.124

2.  Use of Intrathecal Fluorescein in Recurrent Meningitis after Cochlear Implantation.

Authors:  Swati Tandon; Satinder Singh; Shalabh Sharma; Asish K Lahiri
Journal:  Iran J Otorhinolaryngol       Date:  2016-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.