Literature DB >> 9001602

Improvement of the positional accuracy of EIT images of the head using a Lagrange multiplier reconstruction algorithm with diametric excitation.

R H Bayford1, K G Boone, Y Hanquan, D S Holder.   

Abstract

A novel algorithm for the reconstruction of dynamic images using diametric excitation has been developed. The algorithm is specifically designed to image impedance changes in the brain using boundary data obtained from scalp electrodes by incorporating a priori information. The a priori information is obtain by solving the forward problem using a finite-element model (FEM) which includes the discontinuity of the skull resistivity. The advantages with this new approach are that the sensitivity and accuracy of the location of the impedance changes are improved compared to methods based on adjacent excitation.

Entities:  

Mesh:

Year:  1996        PMID: 9001602     DOI: 10.1088/0967-3334/17/4a/008

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  2 in total

1.  Intracranial electrical impedance tomography: a method of continuous monitoring in an animal model of head trauma.

Authors:  Preston K Manwaring; Karen L Moodie; Alexander Hartov; Kim H Manwaring; Ryan J Halter
Journal:  Anesth Analg       Date:  2013-07-10       Impact factor: 5.108

2.  Improved Sensing Pulses for Increased Human Head Depth Measurement Sensitivity With Electrical Impedance Spectroscopy.

Authors:  Giorgio Bonmassar; Michael H Lev
Journal:  IEEE Trans Biomed Eng       Date:  2013-09-10       Impact factor: 4.538

  2 in total