| Literature DB >> 8998560 |
S Myint1, W R Daud, A B Mohamad, A A Kadhum.
Abstract
An ethanolic extract of cloves was analyzed by gas chromatography directly to identify eugenol and other major phenolic compounds without previous separation of other components. Separation was performed on a fused-silica capillary column of 30 m x 0.53 mm I.D., 0.53 microns film thickness. The detector was a flame ionization detector. Helium gas at a flow-rate of 3 ml/min was used as a carrier gas. The analysis were performed with linear temperature programming. Nine components were detected and special attention was given to the major phenolic compound, eugenol.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8998560 DOI: 10.1016/0378-4347(96)00008-4
Source DB: PubMed Journal: J Chromatogr B Biomed Appl ISSN: 1572-6495