Literature DB >> 8976930

Value of radiography in diagnosing complications of cardioverter defibrillators implanted without thoracotomy in 437 patients.

A Gupta1, H G Zegel, V S Dravid, S J Nierenberg, D B Freiman.   

Abstract

OBJECTIVE: This study evaluated the usefulness of radiography in assessing the frequency and cause of complications of nonthoracotomy-implanted cardioverter defibrillators.
MATERIALS AND METHODS: Between May 1992 and December 1995, 437 consecutive patients at our institution underwent cardioverter defibrillator placement. Routine follow-up included external device testing at 6 weeks after placement and every 3 months thereafter. Chest radiographs were obtained immediately after placement, annually, and at the time of any suspected complication. Retrospective review of radiographs and medical charts was done for all patients with clinical complications.
RESULTS: Forty-five complications (10%) were clinically diagnosed: lead or patch fracture in fifteen (33%) patients, electric lead dysfunction in eight (18%), infection in eight (18%), lead retraction in six (13%), patch fold in two (5%), hematoma in two (5%), and other complications in four (9%) patients. Eighteen complications (40%) were radiographically evident. Lead retraction, hematoma, patch fold, patch migration, and the twiddler syndrome were radiographically confirmed in 100% of cases. The average time for these complications to be detected was 68 days; 92% were detected within 23 days. Conversely, only four (27%) lead fractures, one (13%) electric lead dysfunction, and one (13%) infection were radiographically confirmed. These latter complications were discovered an average of 579 days after cardioverter defibrillator placement.
CONCLUSION: Radiography plays a secondary role in the diagnosis of cardioverter defibrillator complications and is particularly limited beyond 1 month after placement. Radiographs may be helpful in the first month after placement because early complications are the most radiographically apparent.

Entities:  

Mesh:

Year:  1997        PMID: 8976930     DOI: 10.2214/ajr.168.1.8976930

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  [Transient unilateral diaphragmatic myoclonus. Life-threatening twiddler's syndrome in a patient with an automatic implantable cardioverter-defibrillator].

Authors:  M Schorl; R Engling
Journal:  Nervenarzt       Date:  2010-12       Impact factor: 1.214

2.  Endovascular extraction techniques: Part 3: Results and indications in patients with an ICD.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-06       Impact factor: 2.380

3.  Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis.

Authors:  Chunliu Zhan; William B Baine; Artyom Sedrakyan; Claudia Steiner
Journal:  J Gen Intern Med       Date:  2008-01       Impact factor: 5.128

  3 in total