Literature DB >> 8973953

Input-output nonlinearities and time delays increase tracking errors in hand grasp neuroprostheses.

M M Adamczyk1, P E Crago.   

Abstract

Tracking tasks were designed to study how different input-output relationships (nonlinearities and time delays) would affect the performance of functional neuromuscular stimulation (FNS) hand grasp neuroprostheses. Simulated hand grasp neuroprostheses and real hand grasp neuroprostheses with and without closed-loop control were evaluated, with the subjects adjusting the input to the system so that the output would match a visual target track at three different bandwidths (able bodied subjects with the simulated systems and neuroprosthesis users with the real systems). For both systems, the tracking error increased as the input-output nonlinearity increased. Other factors that affected tracking performance were the target bandwidth and delays in the neuroprosthesis. The results support the hypothesis that hand grasp neuroprostheses with linear input-output properties and no delays will be controlled more accurately.

Mesh:

Year:  1996        PMID: 8973953     DOI: 10.1109/86.547927

Source DB:  PubMed          Journal:  IEEE Trans Rehabil Eng        ISSN: 1063-6528


  1 in total

1.  Effectiveness of supplemental grasp-force feedback in the presence of vision.

Authors:  M Zafar; C L Van Doren
Journal:  Med Biol Eng Comput       Date:  2000-05       Impact factor: 3.079

  1 in total