Reticulocyte quantification by Coulter MAXM VCS (volume, conductivity, light scatter) technology.

M Buttarello1, P Bulian, M D Prà, P Barbera, P Rizzotti.   

Abstract

In this study the ability of the Coulter MAXM analyzer to quantify reticulocytes was evaluated. The results were compared with those obtained by a microscopic method according to NCCLS H44-P recommendations and with the results from the automated analyzer Sysmex R-1000. Duplicate samples from 330 patients were analyzed. The reference intervals obtained with the three methods were: MAXM 0.37-1.80%, median 0.83%; manual 0.40-2.30%, median 1.00%; R-1000 0.60-1.95%, median 1.06%. The imprecision (CV) at all concentrations is lower than the microscopic method (low 16.1% vs 67%; normal 16.9% vs 28.9%; high 9.5% vs 13.0%). The MAXM shows a good overall correlation with the microscopic method (intercept 0.01, slope 0.89, r2 = 0.87) despite evidence of a significant overestimation at low concentrations [difference (D) 0.30] and a moderate underestimation at normal (D -0.15) and high (D -1.04) concentrations; the same behavior is shown in comparison with results from the R-1000, with an overall underestimation by MAXM (D -0.26). When compared with the microscopic method, MAXM shows a modest sensitivity at low reticulocyte counts (54.8%) and satisfactory sensitivity for high counts (87.3%), with an overall agreement of 81.3%.

Entities:  

Mesh:

Year:  1996        PMID: 8969628

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

Review 1.  Reticulocytes and reticulocyte enumeration.

Authors:  R S Riley; J M Ben-Ezra; R Goel; A Tidwell
Journal:  J Clin Lab Anal       Date:  2001       Impact factor: 2.352

2.  Flow cytometric reticulocyte counting: a comparison between two methods.

Authors:  M Maconi; P Danise; L Cavalca; D Formisano
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

3.  Comparison Between Manual and Automated Methods of Counting Reticulocytes and the Effect of Sample Storage on Reticulocyte Count: A Cross-Sectional Study from Southern India.

Authors:  Linet George; Debdatta Basu; Rakhee Kar
Journal:  Indian J Hematol Blood Transfus       Date:  2021-03-26       Impact factor: 0.900

4.  Dynamics of G6PD activity in patients receiving weekly primaquine for therapy of Plasmodium vivax malaria.

Authors:  Walter R J Taylor; Saorin Kim; Sim Kheng; Sinoun Muth; Pety Tor; Eva Christophel; Mavuto Mukaka; Alexandra Kerleguer; Lucio Luzzatto; J Kevin Baird; Didier Menard
Journal:  PLoS Negl Trop Dis       Date:  2021-09-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.