Literature DB >> 8953153

Characterization of in vivo wear debris from ceramic-ceramic total hip arthroplasties.

S Lerouge1, O Huk, L H Yahia, L Sedel.   

Abstract

In contrast to the much-studied mechanism of aseptic loosening of the metal-polyethylene joint couple, the mechanism responsible for failure of ceramic-ceramic (CC) total hip arthroplasties (THAs) has not been evaluated. The aim of this study was to conduct a systematic characterization of the in vivo wear debris from 15 cases of CC THAs revised for aseptic loosening. Two methods were used to evaluate the wear debris; a semiquantitative histological analysis of H&E-stained periprosthetic pseudomembranes; and an evaluation of isolated debris particles using SEM, energy-dispersive X-ray analysis, and image analysis. The three main types of particulate debris identified were titanium alloy (TiAlV) and alumina ceramic (Al2O3) of prosthetic origin, and zirconium dioxide (ZrO2) from the contrast agent used in the cement for prosthetic fixation. Alumina debris was present in the smallest proportion (12%) and was consistent with the low wear rate of the CC joint couple. Zirconium dioxide debris was present in the greatest proportion (76%) and was an unexpected finding. The ZrO2 debris represented microstructural grains of the original ZrO2 particles added as contrast agent to the cement. The presence of a histiocytic foreign body reaction to ZrO2 debris on histologic sections leads us to believe that these particles play an important role in aseptic loosening of the CC THAs evaluated in this study.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8953153     DOI: 10.1002/(SICI)1097-4636(199612)32:4<627::AID-JBM16>3.0.CO;2-A

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  7 in total

Review 1.  [Implant wear and aseptic loosening. An overview].

Authors:  C Kaddick; I Catelas; P H Pennekamp; M A Wimmer
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

Review 2.  Biological response to prosthetic debris.

Authors:  Diana Bitar; Javad Parvizi
Journal:  World J Orthop       Date:  2015-03-18

3.  Absence of systemic toxicity in mouse model towards BaTiO3 nanoparticulate based eluate treatment.

Authors:  Ashutosh Kumar Dubey; Greeshma Thrivikraman; Bikramjit Basu
Journal:  J Mater Sci Mater Med       Date:  2015-02-06       Impact factor: 3.896

4.  Articulating spacers used in two-stage revision of infected hip and knee prostheses abrade with time.

Authors:  Bernd Fink; Annett Rechtenbach; Hubert Büchner; Sebastian Vogt; Michael Hahn
Journal:  Clin Orthop Relat Res       Date:  2010-07-28       Impact factor: 4.176

5.  Effect of bisphosphonates on the stimulation of macrophages by alumina ceramic particles: a comparison with ultra-high-molecular-weight polyethylene.

Authors:  Alain Petit; Fackson Mwale; John Antoniou; David J Zukor; Olga L Huk
Journal:  J Mater Sci Mater Med       Date:  2006-07       Impact factor: 4.727

6.  Mixed material wear particle isolation from periprosthetic tissue surrounding total joint replacements.

Authors:  Ashley A Stratton-Powell; Sophie Williams; Joanne L Tipper; Anthony C Redmond; Claire L Brockett
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2022-05-09       Impact factor: 3.405

7.  Specific material effects of wear-particle-induced inflammation and osteolysis at the bone-implant interface: A rat model.

Authors:  Lisa K Longhofer; Alexander Chong; Nora M Strong; Paul H Wooley; Shang-You Yang
Journal:  J Orthop Translat       Date:  2016-07-14       Impact factor: 5.191