Strut fracture in De Bakey valve. Successful reoperation and valve replacement.

G L Zumbro, P E Cundey, M E Fishback, R F Galloway.   

Abstract

Strut fracture in a De Bakey aortic valve is reported. The theoretical disadvantage of a Pyrolite carbon--coated ball and titanium struts is mentioned. Periodic cinefluoroscopy is recommended in all patients with a De Bakey aortic valve to avoid the complications associated with strut wear or fracture.

Entities:  

Mesh:

Year:  1977        PMID: 895182

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  1 in total

1.  Asymptomatic strut fracture in DeBakey-Surgitool aortic valves: incidence, management, and metallurgic aspects.

Authors:  J Von Der Emde J; U Eberlein; J Breme J
Journal:  Tex Heart Inst J       Date:  1990