Literature DB >> 8950652

Fluid dynamics of a partially collapsible stenosis in a flow model of the coronary circulation.

M Siebes1, C S Campbell, D Z D'Argenio.   

Abstract

The influence of passive vasomotion on the pressure drop-flow (delta P-Q) characteristics of a partially compliant stenosis was studied in an in vitro model of the coronary circulation. Twelve stenosis models of different severities (50 to 90 percent area reduction) and degrees of flexible wall (0 to 1/2 of the wall circumference) were inserted into thin-walled latex tubing and pressure and flow data were collected during simulated cardiac cycles. In general, the pressure drop increased with increasing fraction of flexible wall for a given flow rate and stenosis severity. The magnitude of this effect was directly dependent upon the underlying stenosis severity. The diastolic delta P-Q relationship of severe, compliant models exhibited features of partial collapse with an increase in pressure drop at a decreasing flow rate. It is concluded that passive vasomotion of a normal wall segment at an eccentric stenosis in response to periodic changes in intraluminal pressure causes dimensional changes in the residual lumen area which can strongly affect the hemodynamic characteristics of the stenosis during the cardiac cycle. This mechanism may have important implications for the onset of plaque fracture and the prediction of the functional significance of a coronary stenosis based on quantitative angiogram analysis.

Entities:  

Mesh:

Year:  1996        PMID: 8950652     DOI: 10.1115/1.2796035

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  A case report: adenosine triggered myocardial infarction during myocardial perfusion stress test imaging in a diabetic patient.

Authors:  Maarten de Mulder; Menno van Gameren; Eric A van Asperen; Martijn Meuwissen
Journal:  Eur Heart J Case Rep       Date:  2021-04-30

2.  Improved Functional Assessment of Ischemic Severity Using 3D Printed Models.

Authors:  Kranthi K Kolli; Sun-Joo Jang; Abdul Zahid; Alexandre Caprio; Seyedhamidreza Alaie; Amir Ali Amiri Moghadam; Patricia Xu; Robert Shepherd; Bobak Mosadegh; Simon Dunham
Journal:  Front Cardiovasc Med       Date:  2022-06-30

3.  Effect of Varying Hemodynamic and Vascular Conditions on Fractional Flow Reserve: An In Vitro Study.

Authors:  Kranthi K Kolli; James K Min; Seongmin Ha; Hilary Soohoo; Guanglei Xiong
Journal:  J Am Heart Assoc       Date:  2016-06-30       Impact factor: 5.501