Literature DB >> 8947905

A method for verifying treatment times for simple high-dose-rate endobronchial brachytherapy procedures.

A V Miller1, M G Davis, J L Horton.   

Abstract

An empirical method for verifying the total treatment time for either a one- or a two-catheter high-dose-rate procedure has been developed. The method can be performed quickly and allows for easy verification of the accuracy of the treatment time arrived at by a computerized planning system. The method is designed to confirm the treatment time to within 10%.

Mesh:

Year:  1996        PMID: 8947905     DOI: 10.1118/1.597753

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  A method for checking high dose rate treatment times for vaginal applicators.

Authors:  C S Mayo; K Ulin
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

  1 in total