Literature DB >> 8946672

External quality assessment of Syva Emit and Abbott TDx II assays for methotrexate in serum.

J F Wilson1, L M Tsanaclis, K Barnett.   

Abstract

The Syva Emit and Abbott TDx II kits for determination of methotrexate in serum were compared using data from 14 samples distributed by the United Kingdom National External Quality Assessment Scheme to a mean of 38 European laboratories. For methotrexate concentrations above 0.2 mumol/L, there was no significant difference in the bias or coefficient of variation of measurements between the two techniques. A significantly greater number of measurements by Syva Emit (6.9%) were rejected as outliers > 3 SD from the sample mean compared with Abbott TDx (1.3%). The lower sensitivity of the Syva Emit assay was evident in data reported for samples containing methotrexate concentrations below 0.2 mumol/L.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8946672     DOI: 10.1097/00007691-199612000-00016

Source DB:  PubMed          Journal:  Ther Drug Monit        ISSN: 0163-4356            Impact factor:   3.681


  1 in total

1.  Establishing Syva Methotrexate Assay on Siemens Dimension RxL Analyzer: Experience in a Tertiary Cancer Care Laboratory.

Authors:  Preeti Chavan; Vivek Bhat; Ashwadeep Karmore; Rajani Mohite; Swati Waykar; Hemali Kadu
Journal:  J Lab Physicians       Date:  2017 Jan-Mar