Geometric design improvements for femoral graft-artery junctions mitigating restenosis.

M Lei1, C Kleinstreuer, J P Archie.   

Abstract

The present study is based on the hypothesis that nonuniform hemodynamics, represented by large time-averaged wall shear stress gradients, trigger abnormal biological processes leading to rapid restenosis, i.e. excessive tissue overgrowth and renewed plaque formation, and hence early graft failure. It implies that this problem may be significantly mitigated by finding graft-artery bypass configurations for which the wall shear stress gradient is approximately zero and hence nearly uniform hemodynamics is achieved. These fluid flow and geometric design considerations are applied to four different end-to-side anastomoses for the distal end of a femoral artery bypass with an appropriate test input pulse and a typical 20-80 flow division. A validated finite-volume code has been used to compute the transient three-dimensional velocity vector fields, wall shear stress distributions and surface contours of the wall shear stress gradients. It is shown that large anastomotic flow areas, small continuously changing bifurcation angles, and smooth junction wall curvatures reduce local time-averaged wall shear stress gradients significantly and hence should mitigate restenosis.

Entities:  

Mesh:

Year:  1996        PMID: 8945659

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  6 in total

1.  Modelling medical devices: the application of bioengineering in surgery.

Authors:  P J Prendergast; D E Beverland; A W Blayney; N J Dunne; T F Gorey; P A Grace; B A McCormack; T McGloughlin; P R O'Connell; J F Orr
Journal:  Ir J Med Sci       Date:  1999 Jan-Mar       Impact factor: 1.568

2.  Computer-Aided Patient-Specific Coronary Artery Graft Design Improvements Using CFD Coupled Shape Optimizer.

Authors:  Onur Dur; Sinan Tolga Coskun; Kasim Oguz Coskun; David Frakes; Levent Burak Kara; Kerem Pekkan
Journal:  Cardiovasc Eng Technol       Date:  2010-11-18       Impact factor: 2.495

3.  Numerical Simulation of Physiological Blood Flow in 2-way Coronary Artery Bypass Grafts.

Authors:  Aike Qiao; Youjun Liu; Siyang Li; Hu Zhao
Journal:  J Biol Phys       Date:  2005-05       Impact factor: 1.365

4.  Evaluation of the hemodynamics in straight 6-mm and tapered 6- to 8-mm grafts as upper arm hemodialysis vascular access.

Authors:  M Sarmast; H Niroomand-Oscuii; F Ghalichi; E Samiei
Journal:  Med Biol Eng Comput       Date:  2014-08-12       Impact factor: 2.602

5.  Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening.

Authors:  John F LaDisa; Lars E Olson; Douglas A Hettrick; David C Warltier; Judy R Kersten; Paul S Pagel
Journal:  Biomed Eng Online       Date:  2005-10-26       Impact factor: 2.819

6.  Alterations in regional vascular geometry produced by theoretical stent implantation influence distributions of wall shear stress: analysis of a curved coronary artery using 3D computational fluid dynamics modeling.

Authors:  John F LaDisa; Lars E Olson; Hettrick A Douglas; David C Warltier; Judy R Kersten; Paul S Pagel
Journal:  Biomed Eng Online       Date:  2006-06-16       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.