Hemodialysis access flow measurement. Comparison of ultrasound dilution and duplex ultrasonography.

J Sands1, D Glidden, C Miranda.   

Abstract

Decreased hemodialysis access flow is associated with an increased risk of access thrombosis. Duplex ultrasonography can measure access flow and select a subset of patients at increased risk for access failure. With in-line techniques (ultrasound dilution), access flow can be measured during hemodialysis. This study attempted to determine if access flow measured by ultrasound dilution (QA-T) was comparable to that measured by duplex ultrasonography (QA-S). The authors performed 66 simultaneous measurements of hemodialysis access flow in 19 patients by ultrasound dilution and duplex ultrasound with time-domain correlation during 19 hemodialysis treatments. The mean access flow was 1,086 +/- 505 ml/min by ultrasound dilution and 1,026 +/- 513 ml/min with duplex ultrasonography (NS). Regression analysis revealed a linear relationship between the two techniques described by the equation QAT = 246.14 + 0.8104(QAS) (correlation coefficient of 0.83; p < 0.0001). Measurement of hemodialysis access flow by ultrasound dilution was essentially equivalent to that obtained by duplex ultrasound. Additional studies are needed to determine if regular in-line flow measurements can predict and prevent future access thrombosis and decreased the cost of access management.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8945015

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  3 in total

1.  Ultrasound dilution and thermodilution versus color Doppler ultrasound for arteriovenous fistula assessment in children on hemodialysis.

Authors:  Vasiliki Karava; Theresa Kwon; Gilbert Franco; Deschenes Georges; Marie-Alice Macher; Julien Hogan
Journal:  Pediatr Nephrol       Date:  2019-07-19       Impact factor: 3.714

2.  Surveillance of hemodialysis vascular access.

Authors:  William L Whittier
Journal:  Semin Intervent Radiol       Date:  2009-06       Impact factor: 1.513

Review 3.  Hemodialysis vascular access monitoring: current concepts.

Authors:  Michael Allon; Michelle L Robbin
Journal:  Hemodial Int       Date:  2009-04       Impact factor: 1.812

  3 in total