| Literature DB >> 8934311 |
F X Diamond1, W E Vickery, J de Kanel.
Abstract
The Zymark RapidTrace automated solid-phase extraction system has been evaluated for use in the urine drug-testing laboratory. Methods for cocaine metabolite (benzoylecgonine) and opiates (codeine and morphine) are described and evaluated. The linearity, precision, accuracy, recovery, and carryover statistics of these analytical protocols are presented. The advantages of extraction using the RapidTrace instead of solid-phase extraction with the manual vacuum manifold or liquid-liquid extraction methods are described.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8934311 DOI: 10.1093/jat/20.7.587
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367