| Literature DB >> 8933975 |
D Shim1, D S Wechsler, T R Lloyd, R H Beekman.
Abstract
We describe the development of hemolysis from moderate residual shunting across a patent ductus arteriosus following coil embolization. The fall in hemoglobin levels from 11.6 to 6.0 gm/dl necessitated a second coil procedure which resulted in complete closure of the residual shunting and resolution of hemolysis. Therefore, appearance of anemia following coil embolization of patent ductus arteriosus should be monitored closely; however, repeat coil embolization with elimination of residual shunt will lead to prompt recovery of normal hemoglobin levels.Entities:
Mesh:
Year: 1996 PMID: 8933975 DOI: 10.1002/(SICI)1097-0304(199611)39:3<287::AID-CCD17>3.0.CO;2-C
Source DB: PubMed Journal: Cathet Cardiovasc Diagn ISSN: 0098-6569