Determination of reference method values by isotope dilution-gas chromatography/mass spectrometry: a five years' experience of two European Reference Laboratories.

L M Thienpont1, B Van Nieuwenhove, D Stöckl, H Reinauer, A P De Leenheer.   

Abstract

We report on the cooperation of two European Reference Laboratories for the determination of reference method values in serum based materials intended for use in internal accuracy control and external quality assessment. Reference method values were determined by isotope dilution-gas chromatography/mass spectrometry for aldosterone, cortisol, oestradiol-17 beta, progesterone, testosterone, thyroxine, theophylline, cholesterol, creatinine, glucose, total triacylglycerols and uric acid. All determinations were done in parallel in the two laboratories, independently and within certain time constraints. The general measurement design consisted of duplicate measurement of each sample on three different occasions. In each laboratory, rigorous internal quality control was performed according to predefined analytical quality specifications. This was done using certified reference materials. If not available, control materials targeted before by the two laboratories were utilized. Here we present the results of the cooperation during five years. We discuss the precision and accuracy achieved, the between-laboratory agreement and the total analytical error. For the hormones and theophylline, the mean overall coefficient of variation for both laboratories (calculated from the measurements on three days) was always < 2%, for the substrates < 1%. For all substances, the method bias (estimated from several measurement series over the five years) was < 1%, and the average deviation of the results between the two laboratories was < 1.2%. The maximum total analytical error was in all cases < 3%. These data demonstrate that current reference methodology is able to guarantee a stable level of high quality of performance, and that reference laboratories of today are capable of providing, in due time, adequate service in the framework of accuracy-based harmonization of methods in routine laboratory medicine.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8933112

Source DB:  PubMed          Journal:  Eur J Clin Chem Clin Biochem        ISSN: 0939-4974


  6 in total

1.  Accuracy-based proficiency testing for testosterone measurements with immunoassays and liquid chromatography-mass spectrometry.

Authors:  Zhimin Tim Cao; Julianne Cook Botelho; Robert Rej; Hubert Vesper
Journal:  Clin Chim Acta       Date:  2017-03-10       Impact factor: 3.786

2.  Steroid hormone levels in pregnancy and 1 year postpartum using isotope dilution tandem mass spectrometry.

Authors:  Offie P Soldin; Tiedong Guo; Elisabete Weiderpass; Rochelle E Tractenberg; Leena Hilakivi-Clarke; Steven J Soldin
Journal:  Fertil Steril       Date:  2005-09       Impact factor: 7.329

3.  Determination of uric acid in biological samples by high performance liquid chromatography-electrospray ionization-tandem mass spectrometry and study on pathogenesis of pulmonary arterial hypertension in pulmonary artery endothelium cells.

Authors:  Qiaozhi Li; Yanli Qiu; Weina Han; Yaqin Zheng; Xiaoying Wang; Dandan Xiao; Min Mao; Qian Li
Journal:  RSC Adv       Date:  2018-07-18       Impact factor: 4.036

4.  Revisiting hyper- and hypo-androgenism by tandem mass spectrometry.

Authors:  Flaminia Fanelli; Alessandra Gambineri; Marco Mezzullo; Valentina Vicennati; Carla Pelusi; Renato Pasquali; Uberto Pagotto
Journal:  Rev Endocr Metab Disord       Date:  2013-06       Impact factor: 6.514

5.  A sensitive and specific liquid chromatography-tandem mass spectrometry method for the determination of intracellular and extracellular uric acid.

Authors:  Kyung Mee Kim; George N Henderson; Xiaosen Ouyang; Reginald F Frye; Yuri Y Sautin; Daniel I Feig; Richard J Johnson
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2009-05-27       Impact factor: 3.205

6.  Validated low-volume aldosterone immunoassay tailored to GCLP-compliant investigations in small sample volumes.

Authors:  J Schaefer; B B Burckhardt; J Tins; A Bartel; S Laeer
Journal:  Pract Lab Med       Date:  2017-07-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.