Nonuniform heating patterns of commercial electrodes for radiofrequency catheter ablation.

M S Mirotznik1, D Schwartzman.   

Abstract

INTRODUCTION: Previous investigations of the biophysics of myocardial ablation using radiofrequency energy have focused mainly on the electrode-myocardial interface. Relatively little has been written about the overall function of commercial electrodes used for ablation. The purpose of this study was to evaluate the heating patterns of several commercial electrodes used for myocardial ablation. METHODS AND
RESULTS: The specific absorption rate, a unit of measure proportional to the initial rate of rise of the temperature at the electrode-tissue interface and reflecting the energy deposition from the electrode, was measured during radiofrequency energy application at multiple points around each electrode. These measurements were combined to form a specific absorption rate pattern, which reflected the spatial pattern of energy deposition from the electrode. There was significant nonuniformity of the specific absorption rate pattern of each electrode evaluated. The proximal edge of the electrode and the curvature of the distal portion of the electrode were two sites of accentuated heating relative to the remaining portions of the electrode. The magnitude of this nonuniformity was similar between the 4-mm length electrodes of different manufacturers, but increased progressively in larger (5- and 8-mm length) electrodes.
CONCLUSIONS: The heating patterns of commercial electrodes used for myocardial ablation are nonuniform. These data have implications for electrode design and utilization.

Mesh:

Year:  1996        PMID: 8930737     DOI: 10.1111/j.1540-8167.1996.tb00481.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  3 in total

1.  Comparison of irrigated electrode designs for radiofrequency ablation of myocardium.

Authors:  D Demazumder; M S Mirotznik; D Schwartzman
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

2.  Ring electrode for radio-frequency heating of the cornea: modelling and in vitro experiments.

Authors:  E J Berjano; J Saiz; J L Alió; J M Ferrero
Journal:  Med Biol Eng Comput       Date:  2003-11       Impact factor: 2.602

3.  Biophysics of radiofrequency ablation using an irrigated electrode.

Authors:  D Demazumder; M S Mirotznik; D Schwartzman
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.