| Literature DB >> 8906078 |
M L'Hermite-Balériaux1, G Copinschi, E Van Cauter.
Abstract
We compared the data from four growth hormone (GH) immunoassays for analyzing 24-h GH profiles in four apparently normal subjects and four obese subjects (508 serum samples). The detection limit was 0.02 microgram/L for one immunochemiluminometric assay (ICMA), 0.1 microgram/L for two IRMAs, and 0.4 microgram/L for one RIA. All GH pulses with a peak ICMA value > 1 microgram/L were detected by each of the other methods. Overall, the correlation coefficient between the values obtained with all four assays exceeded 0.90. However, for GH concentrations < or = 0.25 microgram/L, acceptable concordance (r2 > or = 0.80) was reached only between the ICMA and one IRMA; between the ICMA and the RIA, concordance was acceptable only for GH concentrations > or = 10 micrograms/L. In the normal subjects, the percentage of undetectable values was 0% with the ICMA but 29% with one of the IRMAs; in obese subjects, the corresponding values were 12% and 38%.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8906078
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327