| Literature DB >> 8896324 |
Abstract
The development of a microgravimetric immunobiosensor using a piezoelectric quartz crystal as a detector requires a stable and reproducible immobilization method for ligand binding. The method of silanization using 3-aminopropyltriethoxysilane (APTES) has been widely used for activating the carrier surface. In the present study, APTES deposition on a piezoelectric crystal surface was studied under various solvent conditions. A fluorescence method, using fluorescence isothiocyanate as a dye, was demonstrated for the quantification of amino groups on the silanized piezoelectric crystal surface. The optimum binding conditions of APTES deposition on a piezoelectric crystal surface were incorporated for the covalent immobilization of protein on the crystal surface in developing a stable and sensitive microgravimetric immunobiosensor. Determination of immunoglobulin G (IgG) concentration was performed using APTES modified piezoelectric crystals coated with protein G. The resonant frequency shift, resulting from the formation of protein G-IgG complex on the crystal surface, correlated with the concentration of IgG in the range 10 ng/ml to 0.1 mg/ml. The APTES modified, protein G coated crystal were found to be quite stable and did not show a significant loss of sensitivity even after 12 weeks of storage at 4 degrees C in a desiccator.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8896324 DOI: 10.1016/0956-5663(96)88084-5
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618