| Literature DB >> 8891444 |
Abstract
Individual synthetic vesicles 0.1-1.0 micron in diameter are sized by using single-particle fluorescence emission spectra obtained in a custom sheath flow cell with an imaging spectrograph and a charge-coupled device. Data are acquired at 1 Hz, with limits of detection (3 sigma) less than 6.0 x 10(3) and 1.0 x 10(4) molecules of free sulforhodamine 101 and fluorescein, respectively, and with a spectral range for fluorescence emission collection from 350 to 800 nm (0.45 nm/pixel resolution). The system is used for small-particle population analysis by analyzing a suspension of submicron, unilamellar, synthetic vesicles prepared by standard procedures with phosphatidylserine and Texas red- or fluorescein head-group-conjugated dihydropalmitoylphosphatidylethanolamine. The submicron particles are individually identified, sized, and discriminated based on single-particle fluorescence emission spectra. Excellent agreement is found between fluorescence sizing data and transmission electron microscopic measurements.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8891444 DOI: 10.1002/(SICI)1097-0320(19961001)25:2<144::AID-CYTO3>3.0.CO;2-H
Source DB: PubMed Journal: Cytometry ISSN: 0196-4763