Abstract
Pedicle screws are commonly used in spinal reconstruction, and failure of pedicle screws due to bending is a significant clinical problem. To measure the moments typically placed on pedicle screws in situ we instrumented 7 mm Cotrel-Dubousset (CD) pedicle screws with internally mounted strain gauges. The screws were designed to measure flexion-extension moments at a single cross-section as dictated by strain gauge placement. It is possible to measure moments of up to 12 Nm at any location along the length of the screw by constructing transducers with varying strain gauge placements. These transducers are capable of measuring moments at points located within the vertebra including the pedicle, which is where failure usually occurs clinically. Transducer output was both linear and reproducible. These transducers are being used to investigate the load transfer characteristics between the pedicle screw and the vertebra. This technique could be applied to investigations of load sharing in reconstruction plates, lag-screws, and cross-locked intramedullary nails.Mesh:
Year: 1996 PMID: 8872267 DOI: 10.1115/1.2796027
Source DB: PubMed Journal: J Biomech Eng ISSN: 0148-0731 Impact factor: 2.097