Literature DB >> 8865420

Simultaneous measurement of total hemoglobin and its derivatives in blood using CO-oximeters: analytical principles; their application in selecting analytical wavelengths and reference methods; a comparison of the results of the choices made.

J A Brunelle1, A M Degtiarov, R F Moran, L A Race.   

Abstract

Optical methods of quantifying total hemoglobin (tHb), applying the principles of the Lambert-Beer law, have been used both on untreated whole blood and on blood mixed with chemicals to form a stable chromophore, since the earliest days of laboratory medicine. The same principles may be applied for quantitation of the individual hemoglobin derivatives, such as oxyhemoglobin (O2Hb) and deoxyhemoglobin (HHb)1, as well as the non-oxygen transporting "dyshemoglobins", including carboxyhemoglobin (COHb) and methemoglobin (MetHb). The total hemoglobin measurement is typically carried out using a light source with a broad band of visible wavelengths. However, measurement of the derivatives requires using discrete, narrow bands of light in order to differentiate between the small differences of light absorbed by the individual derivatives. Either general-purpose, narrow band-pass spectrophotometers, or special-purpose photometers utilizing a set of fixed wavelengths, commonly referred to as "CO-oximeters" are suitable. Rapid, direct, photometric quantification of the derivatives, necessary in the clinical environment, relies on the specific light absorption characteristics of each hemoglobin derivative at the wavelengths selected, which in turn requires independent and exact knowledge of the concentrations of each entity in reference materials. This report examines the process involved in the selection of wavelengths and reference methods, contrasts the effects of the choices made and discusses some implications and limitations for routine measurement.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8865420     DOI: 10.3109/00365519609088624

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest Suppl        ISSN: 0085-591X


  6 in total

Review 1.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

Review 2.  Clinical chameleons: an emergency medicine focused review of carbon monoxide poisoning.

Authors:  Patrick Chow Ng; Brit Long; Alex Koyfman
Journal:  Intern Emerg Med       Date:  2018-02-12       Impact factor: 3.397

3.  Assessment of Photoplethysmography Method in Extraction of Hemoglobin Concentration.

Authors:  M Azarnoosh; H Doostdar
Journal:  J Biomed Phys Eng       Date:  2019-12-01

4.  Determination of Methemoglobin in Hemoglobin Submicron Particles Using NMR Relaxometry.

Authors:  Waraporn Kaewprayoon; Nittiya Suwannasom; Chiraphat Kloypan; Axel Steffen; Yu Xiong; Eyk Schellenberger; Axel Pruß; Radostina Georgieva; Hans Bäumler
Journal:  Int J Mol Sci       Date:  2020-11-26       Impact factor: 5.923

5.  Unexplained Methemoglobinemia in Coronavirus Disease 2019: A Case Report.

Authors:  Hina Faisal; Alexi Bloom; A Osama Gaber
Journal:  A A Pract       Date:  2020-07

6.  Measuring hemoglobin spectra: searching for carbamino-hemoglobin.

Authors:  Emmanuel Dervieux; Quentin Bodinier; Wilfried Uhring; Michaël Théron
Journal:  J Biomed Opt       Date:  2020-10       Impact factor: 3.170