Supraorbital artery as an alternative site for oscillometric blood pressure measurement.

T K Lee1, T P Egbert, D R Westenskow.   

Abstract

OBJECTIVE: Noninvasive blood pressure measured from the superficial temporal artery has been shown to correlate well with pressure in the brachial artery. The supraorbital artery may be an even better site for monitoring blood pressure on the forehead because it originates from the internal carotid artery, and it is easier to locate anatomically. This study compares mean pressure measured oscillometrically over the supraorbital artery and at the upper arm.
METHODS: Oscillometric signals from the supraorbital artery were recorded in 20 surgical patients under general anesthesia using a 2.5- x 1-cm bladder attached to the forehead with a self-adhesive pad. Blood pressure was measured simultaneously from the arm using a Dinamap 1846 blood pressure monitor, and the resulting data compared with the supraorbital artery measurements.
RESULTS: The mean difference between 219 pairs of blood pressure measurements, from the forehead and the arm, was 3.8 mm Hg. The standard deviation of the differences was 7.4 mm Hg. The linear regression equation for the data was y = 0.98x + 3.25, with a standard error of estimate of 7.31 mm Hg. The correlation coefficient between the two measurements was 0.82.
CONCLUSIONS: The results show that mean blood pressures measured oscillometrically from the supraorbital and brachial arteries agree and correlate well with each other. The supraorbital artery should be a good alternative site for blood pressure measurement.

Entities:  

Mesh:

Year:  1996        PMID: 8863108     DOI: 10.1007/bf02221749

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

1.  The importance of cuff width in measurement of blood pressure indirectly.

Authors:  L A Geddes; R Tivey
Journal:  Cardiovasc Res Cent Bull       Date:  1976 Jan-Mar

2.  Optimal sites for forehead oscillometric blood pressure monitoring.

Authors:  T K Lee; F H Silva; T P Egbert; D R Westenskow
Journal:  J Clin Monit       Date:  1995-09

3.  Analysis of blood flow through a model of the human arterial system under periodic body acceleration.

Authors:  V K Sud; G S Sekhon
Journal:  J Biomech       Date:  1986       Impact factor: 2.712

4.  Indirect measurement of mean blood pressure in the anesthetized patient.

Authors:  M Yelderman; A K Ream
Journal:  Anesthesiology       Date:  1979-03       Impact factor: 7.892

5.  The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. 1.

Authors:  J A Posey; L A Geddes; H Williams; A G Moore
Journal:  Cardiovasc Res Cent Bull       Date:  1969 Jul-Sep

6.  Noninvasive blood pressure measurement on the temporal artery using the auscultatory method.

Authors:  D Shenoy; W W von Maltzahn; J C Buckey
Journal:  Ann Biomed Eng       Date:  1993 Jul-Aug       Impact factor: 3.934

7.  An assessment of the Dinamap 845.

Authors:  P Hutton; J Dye; C Prys-Roberts
Journal:  Anaesthesia       Date:  1984-03       Impact factor: 6.955

8.  Peroneal nerve palsy in a premature: complication of multiple blood pressure measurements.

Authors:  G P Giacoia
Journal:  Clin Pediatr (Phila)       Date:  1981-09       Impact factor: 1.168

9.  The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure--part ii.

Authors:  G W Mauck; C R Smith; L A Geddes; J D Bourland
Journal:  J Biomech Eng       Date:  1980-02       Impact factor: 2.097

10.  Measurements of base blood pressure during sleep and its clinical significance in hypertensive patients.

Authors:  O Tochikubo; H Ochiai; T Oota; E Miyajima; Y Kaneko
Journal:  Jpn Circ J       Date:  1987-10
View more
  1 in total

1.  Comparison of blood pressure measured by oscillometry from the supraorbital artery and invasively from the radial artery.

Authors:  T K Lee; D R Westenkow
Journal:  J Clin Monit Comput       Date:  1998-02       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.