Literature DB >> 8848923

No additional metal particle formation using the needle-through-needle combined epidural/spinal technique.

N Herman1, J Molin, K G Knape.   

Abstract

BACKGROUND: Combined epidural/spinal analgesia utilizing a needle-through-needle technique has become very popular in anesthesia. However, findings of concave deformities at the orifice of Tuohy needles after spinal needle passage have raised concerns that metal fragments might be deposited within the epidural space. This study was proposed to investigate whether the needle-through-needle technique does produce metallic flecks.
METHODS: Ten unused Tuohy and Hustead epidural needles were inspected microscopically and photomicrographed prior to flushing saline through each into a single tissue culture well. After drying, a single pass was made with a 120 mm 24-gauge Sprotte needle through each epidural needle to maximal extension while the orifice was within another tissue culture well. Each needle was again flushed into a third well before reexamination and photomicroscopy. Each of the wells was inspected for metallic particles by microscopy. Additional freshly unpackaged Tuohy needles were microscopically examined after exposure to a magnetic field.
RESULTS: Comparison of micrographs before and after needle experiments revealed concave deformities at the orifice of all the Tuohy and Hustead needles. No particles were observed in either of the two saline-flushed wells or within the well in which the needle-through-needle passes were made. Inspection of unused Tuohy needles exposed to a magnetic field revealed metal filings "standing up" along the bevel of every needle examined.
CONCLUSIONS: Metallic particles are not produced by the needle-through-needle technique. However, metal particles are an apparent contaminant of all epidural needles and are probably routinely introduced into patients when the needle is placed.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8848923     DOI: 10.1111/j.1399-6576.1996.tb04424.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  5 in total

1.  [Structural damage of epidural catheters during combined spinal epidural anaesthesia: a lightmicroscopical study].

Authors:  B M Wahlen; M David; A Stanek
Journal:  Anaesthesist       Date:  2006-05       Impact factor: 1.041

2.  Needle-through-needle compared to Eldor needle technique.

Authors:  J Eldor; S Levine
Journal:  J Anesth       Date:  1997-06       Impact factor: 2.078

3.  Aseptic meningitis using the needle-through-needle technique.

Authors:  J Eldor
Journal:  Can J Anaesth       Date:  1996-11       Impact factor: 5.063

Review 4.  Complications of regional anaesthesia Incidence and prevention.

Authors:  K A Faccenda; B T Finucane
Journal:  Drug Saf       Date:  2001       Impact factor: 5.606

Review 5.  Regional anaesthesia in pre-eclampsia: advantages and disadvantages.

Authors:  Nanda Gopal Mandal; Sridhar Surapaneni
Journal:  Drugs       Date:  2004       Impact factor: 9.546

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.