Phantom study of chest radiography with storage phosphor, selenium, and film-screen systems.

M Kehler1, K Lyttkens, B Andersson, P Hochbergs, C G Lindberg, J Medin, A J Nordström, J Sanfridsson, J Vojciechowski.   

Abstract

PURPOSE: Chest radiographs from 3 digital systems--2 based on luminescent phosphors and one on selenium--and a conventional film-screen system were evaluated and compared.
MATERIAL AND METHODS: Computed radiography (CR) has for the past years been dominated by a single manufacturer, but now several systems have been marketed. Using a chest phantom and, as test objects, 2 simulated tumours for the lung and mediastinum, respectively, and one object simulating pulmonary lines, a total of 400 exposures were made, 100 on each system. The test objects were placed randomly with a ratio of presence/absence of each object of about 50. Six radiologists participated, 2 residents and 4 staff members. A receiver operating characteristics (ROC) analysis was performed with construction of curves, and the difference between the curves was estimated with a 2-tailed paired t-test. RESULTS AND
CONCLUSION: The selenium-based system performed significantly better for pulmonary line detection than all the other systems, and better than one storage phosphor system for the lung "tumour" (p < 0.05), while one storage phosphor system was slightly better than the other in diagnosing all 3 test objects. The score for the film-screen system was only average.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8845264     DOI: 10.1177/02841851960371P170

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  1 in total

1.  Interpretation of normal anatomic structures on chest radiography: comparison of Fuji computed radiography (FCR) 5501D with FCR 5000 and screen-film system.

Authors:  Kazuaki Nakashima; Kazuto Ashizawa; Makoto Ochi; Rashid Hashmi; Kuniaki Hayashi; Shinichi Gotoh; Sumihisa Honda; Akito Igarashi; Takao Komaki
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.