Literature DB >> 8834723

Tears in bioprosthetic heart valve leaflets without calcific degeneration.

F Haziza1, G Papouin, B Barratt-Boyes, G Christie, R Whitlock.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The mechanism of structural failure of bioprosthetic valves is still not clearly understood. This study was undertaken to assess pure leaflet tear as a mode of failure in porcine and pericardial bioprostheses.
METHODS: Of 246 bioprosthetic valves (109 porcine, 137 pericardial) implanted between 1975 and 1991, 101 had to be explanted and served as the study population.
RESULTS: The reasons for valve failure were calcific degeneration in 73, pure tear in 12, and endocarditis in 10. Six other patients had a perivalvar leak. The mean age at operation was 32 years. Freedom from degeneration at 10 years was 45 +/- 7% and from pure tear it was 92 +/- 2%. The hazard functions were strikingly different, as that for degeneration showed a progressive increase while that for pure tear peaked at six years post-implant. The mean age of the patients with pure tear was 41 years and for degeneration it was 24 years (p = 0.00001). The reasons for the difference in hazard function are discussed. The characteristic clinical features of pure tear allow clinical diagnosis in the majority of patients.
CONCLUSION: Pure tear is the result of uneven tissue loading with tearing occurring at sites of maximal stress. The four possible mechanisms in pericardial valves are (a) intense stress concentration at the top of the stent post (commissure); (b) compression stress below the top of the post; (c) abrasion stress in tissue mounted outside the frame; and (d) increased bending stresses on leaflet opening. In stent-mounted porcine valves, pure tear is related to incorrect mounting or to increased bending stresses.

Entities:  

Mesh:

Year:  1996        PMID: 8834723

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  9 in total

1.  Comparative study of the mechanical behaviour of a cyanoacrylate and a bioadhesive.

Authors:  J M García Páez; E Jorge Herrero; A Rocha; M Maestro; J L Castillo-Olivares; I Millan; A Carrera Sanmartin; A Cordon
Journal:  J Mater Sci Mater Med       Date:  2004-02       Impact factor: 3.896

2.  A new method for selecting calf pericardium for use in cardiac bioprostheses on the basis of morphological and mechanical criteria.

Authors:  J M García Páez; E Jorge-Herrero; A Carrera; I Millán; A Rocha; A Cordón; J Salvador; N Sainz; J Méndez; J L Castillo-Olivares
Journal:  J Mater Sci Mater Med       Date:  2001-08       Impact factor: 3.896

3.  Porcine pericardial membrane subjected to tensile testing: preliminary study of the process of selecting tissue for use in the construction of cardiac bioprostheses.

Authors:  J M Garcia Páez; E Jorge-Herrero; A Carrera; I Millán; A Rocha; J Salvador; J Mendez; G Téllez; J L Castillo-Olivares
Journal:  J Mater Sci Mater Med       Date:  2001-05       Impact factor: 3.896

Review 4.  Pure Cusp Tear of Trifecta Bioprosthesis 2 Years after Aortic Valve Replacement.

Authors:  Masaki Hamamoto; Taira Kobayashi; Masamichi Ozawa; Kosuke Yoshimura
Journal:  Ann Thorac Cardiovasc Surg       Date:  2016-12-14       Impact factor: 1.520

5.  Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties.

Authors:  Caitlin Martin; Wei Sun
Journal:  Biomech Model Mechanobiol       Date:  2013-10-04

6.  Form Follows Function: Advances in Trilayered Structure Replication for Aortic Heart Valve Tissue Engineering.

Authors:  Dan T Simionescu; Joseph Chen; Michael Jaeggli; Bo Wang; Jun Liao
Journal:  J Healthc Eng       Date:  2012-06       Impact factor: 2.682

7.  Mechanical effects of increases in the load applied in uniaxial and biaxial tensile testing: Part I. Calf pericardium.

Authors:  J M García Páez; E Jorge; A Rocha; M Maestro; J L Castillo-Olivares; I Millan; A Carrera; A Cordon; G Tellez; R Burgos
Journal:  J Mater Sci Mater Med       Date:  2002-04       Impact factor: 3.896

Review 8.  Fatigue damage of collagenous tissues: experiment, modeling and simulation studies.

Authors:  Caitlin Martin; Wei Sun
Journal:  J Long Term Eff Med Implants       Date:  2015

9.  Is cell regeneration and infiltration a double edged sword for porcine aortic valve deterioration? A large cohort of histopathological analysis.

Authors:  Li Li; Xuejing Duan; Hongyue Wang; Yang Sun; Wei Zhao; Yang Lu; Hongyu Xu; Yiwei You; Qingzhi Wang
Journal:  BMC Cardiovasc Disord       Date:  2022-07-28       Impact factor: 2.174

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.