| Literature DB >> 8832427 |
P O Erah1, D A Barrett, P N Shaw.
Abstract
A simple and selective ion-pair HPLC method has been developed for the analysis of clarithromycin in aqueous solutions and in gastric juice. A Hypersil ODS 5-microns (150 x 4.6 mm I.D.) column was used with a mobile phase consisting of acetonitrile-aqueous 0.05 M phosphate buffer (pH 4.6) containing 5 mM l-octanesulphonic acid (50:50, v/v). The column temperature was 50 degrees C and detection was by UV absorption (210 nm). The limits of detection of 50-microliters samples were 0.4 microgram/ml (aqueous) and 0.78 microgram/ml (0.5 ml gastric juice) or better. The assay was linear in the range of 1.56 to 100 micrograms/ml with r2 values greater than 0.99. The recovery from the gastric juice samples was 98.5 +/- 2.9%. The method was applied successfully to determine the stability of clarithromycin in 0.01 M HCl and gastric juice.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8832427 DOI: 10.1016/0378-4347(96)00064-3
Source DB: PubMed Journal: J Chromatogr B Biomed Appl ISSN: 1572-6495