Literature DB >> 8825113

Prevention of catheter-related infections by silver coated central venous catheters in oncological patients.

H Goldschmidt1, U Hahn, H J Salwender, R Haas, B Jansen, P Wolbring, M Rinck, W Hunstein.   

Abstract

Catheter-related infection (CRI) is a serious complication of central venous catheterization. We have investigated the efficacy of a silver-coated polyurethane catheter (Pellethane, Fresenius AG, Germany) in preventing CRI in oncological patients receiving chemotherapy in a phase II study. From November 1992 through April 1994, 266 patients were assigned to receive single lumen catheters, either standard uncoated catheters (UC, n = 113) or silver-coated ones (SC, n = 120). Catheters were inserted into the internal jugular vein after institutional approval and informed consent. Duration of catheterization (UC vs. SC = 13.3 vs. 12.7 days) and leukopenia (< 1.0 x 10(9) WBC/l; 4.3 vs. 3.6 days) were similar in both groups demonstrating a comparable risk for infections. Skin reactions at the catheter entry site were recorded daily. CRI and colonization rates were studied by semiquantitatively culturing intradermal and intravascular segments. CRI were confirmed by blood cultures obtained via catheter and from peripheral veins in cases of suspected sepsis or at the end of catheterization. No adverse effects from the silver-coated catheter could be observed. The bacteriological results showed that SC were colonized (> 15 CFU) in 45.1% and UC in 44.2%. CRI developed in 21.2% of the UC patients but only in 10.2% of the SC patients (p = 0.011). We conclude that this new silver-coated central venous catheter is biocompatible and effective in reducing the incidence of catheter-related infections in oncological patients.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8825113     DOI: 10.1016/s0934-8840(11)80203-3

Source DB:  PubMed          Journal:  Zentralbl Bakteriol        ISSN: 0934-8840


  11 in total

Review 1.  Prevention of infections caused by central venous catheters--established and novel measures.

Authors:  A Bach
Journal:  Infection       Date:  1999       Impact factor: 3.553

Review 2.  Infections associated with medical devices: pathogenesis, management and prophylaxis.

Authors:  Christof von Eiff; Bernd Jansen; Wolfgang Kohnen; Karsten Becker
Journal:  Drugs       Date:  2005       Impact factor: 9.546

Review 3.  Anti-infective-treated central venous catheters: a systematic review of randomized controlled trials.

Authors:  B S Niël-Weise; T Stijnen; P J van den Broek
Journal:  Intensive Care Med       Date:  2007-10-17       Impact factor: 17.440

4.  Sixty day continuous use of subdermal wire electrodes for EEG monitoring during treatment of status epilepticus.

Authors:  Gabriel U Martz; Christina Hucek; Mark Quigg
Journal:  Neurocrit Care       Date:  2009-05-01       Impact factor: 3.210

Review 5.  Silver and Alginates: Role in Wound Healing and Biofilm Control.

Authors:  Steven L Percival; Sara M McCarty
Journal:  Adv Wound Care (New Rochelle)       Date:  2015-07-01       Impact factor: 4.730

6.  Reduced rates of catheter-associated infection by use of a new silver-impregnated central venous catheter.

Authors:  M Böswald; S Lugauer; A Regenfus; G G Braun; P Martus; C Geis; J Scharf; T Bechert; J Greil; J P Guggenbichler
Journal:  Infection       Date:  1999       Impact factor: 3.553

7.  Tolerability and performance of BIP endotracheal tubes with noble metal alloy coating--a randomized clinical evaluation study.

Authors:  Gunilla Björling; Dorota Johansson; Linda Bergström; Shah Jalal; Ivar Kohn; Claes Frostell; Sigridur Kalman
Journal:  BMC Anesthesiol       Date:  2015-12-01       Impact factor: 2.217

8.  Improved ex vivo blood compatibility of central venous catheter with noble metal alloy coating.

Authors:  Manijeh Vafa Homann; Dorota Johansson; Håkan Wallen; Javier Sanchez
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2015-07-06       Impact factor: 3.368

Review 9.  Catheter impregnation, coating or bonding for reducing central venous catheter-related infections in adults.

Authors:  Nai Ming Lai; Nathorn Chaiyakunapruk; Nai An Lai; Elizabeth O'Riordan; Wilson Shu Cheng Pau; Sanjay Saint
Journal:  Cochrane Database Syst Rev       Date:  2016-03-16

10.  Impact of catheter antimicrobial coating on species-specific risk of catheter colonization: a meta-analysis.

Authors:  Aleksey Novikov; Manuel Y Lam; Leonard A Mermel; Anna L Casey; Tom S Elliott; Peter Nightingale
Journal:  Antimicrob Resist Infect Control       Date:  2012-12-03       Impact factor: 4.887

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.