Literature DB >> 8823635

Equivalence of fast flush and square wave testing of blood pressure monitoring systems.

B Kleinman1, S Powell, R M Gardner.   

Abstract

BACKGROUND: The accurate recording of intraarterial pressure depends upon an appropriate dynamic response of the monitoring system. Generation of a square wave (SW) at the catheter tip is the engineering and in vitro laboratory gold standard. Fast flush (FF) testing is the clinical test of choice. Results from these two test methods have been assumed equal but have not been empirically confirmed.
METHODS: We studied three different 5.1 cm catheter sizes (16 G, 18 G, 20 G Becton Dickinson, Sandy, UT) attached to three different lengths of arterial pressure tubing (36 in, 91.4 cm; 72 in, 182.9 cm; 108 in, 274.3 cm). An arterial recording system was assembled in the standard fashion by attaching a catheter to arterial pressure tubing, which was attached to a transducer (TXX-R, Ohmeda, formerly Viggo-Spectramed, Oxnard, CA) whose signal was recorded by a strip chart recorder (Gould 2400, Rolling Meadows, IL). The system was attached to a pressurized saline flush. The catheter tip was inserted into one port of a pressure generator. With the other port of the pressure generator open to atmosphere, FF tests were performed by activating the flush device of the transducer. Subsequent step response signals from the FF tests were then recorded from which natural frequency (fn) and damping coefficient (zeta) were calculated. Next, square waves were generated by closing the port that was open to atmosphere and attaching a signal generator to a pressure generator. Square waves so generated were recorded as described above and natural frequency and damping coefficients calculated. These procedures were repeated after 0.05 cc of air was introduced in the transducer and repeated again in a system containing a damping device (R.O.S.E., Resonant OverShoot Eliminator, Viggo-Spectramed, Oxnard, CA).
RESULTS: There was no significant difference between fn and zeta as calculated from the step response generated from the FF test versus fn and zeta as calculated from the square wave (SW) test in systems without air. However, in systems containing air, fn by FF testing was always less than fn by SW testing for all catheter sizes and extension tubing lengths (p < 0.05). Damping was also always greater by FF testing than by SW testing in systems with air for all catheter sizes and extension tubing lengths (p < 0.05). The R.O.S.E device created marked qualitative differences, although exact fn and zeta could not be quantified.
CONCLUSIONS: For the characterization of dynamic response of invasive blood pressure monitoring systems, the FF test and SW test yield identical results. However, under certain conditions-air, R.O.S.E device-dynamic response as measured by FF testing was not equivalent to dynamic response as measured by the gold standard-the SW test. Specifically, small amounts of air in fluid-filled invasive blood pressure monitoring systems cause a slightly worse dynamic response as measured by FF testing versus the laboratory gold standard-the SW test.

Entities:  

Mesh:

Year:  1996        PMID: 8823635     DOI: 10.1007/bf02078135

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  6 in total

1.  The fast flush test measures the dynamic response of the entire blood pressure monitoring system.

Authors:  B Kleinman; S Powell; P Kumar; R M Gardner
Journal:  Anesthesiology       Date:  1992-12       Impact factor: 7.892

2.  Physiologic recording by modern instruments with particular reference to pressure recording.

Authors:  D L FRY
Journal:  Physiol Rev       Date:  1960-10       Impact factor: 37.312

3.  Catheter-flush system for continuous monitoring of central arterial pulse waveform.

Authors:  R M Gardner; H R Warner; A F Toronto; W D Gaisford
Journal:  J Appl Physiol       Date:  1970-12       Impact factor: 3.531

4.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

5.  Dynamic response of a neonatal catheter-manometer system in situ.

Authors:  H van Langen; P Brienesse; K Kopinga; P Wijn
Journal:  J Clin Monit       Date:  1993-11

6.  System for umbilical artery monitoring.

Authors:  R M Gardner; J Parker; L R Feinauer
Journal:  Crit Care Med       Date:  1982-07       Impact factor: 7.598

  6 in total
  3 in total

1.  Frequency characteristics of pressure transducer kits with inserted pressure-resistant extension tubes.

Authors:  Shigeki Fujiwara; Satoshi Mori; Keiichi Tachihara; Takeshi Yamamoto; Chizuko Yokoe; Uno Imaizumi; Yoshinari Morimoto; Yoichiro Miki; Izumi Toyoguchi; Kazu-Ichi Yoshida; Takeshi Yokoyama
Journal:  J Clin Monit Comput       Date:  2016-03-05       Impact factor: 2.502

2.  The fast flush test--is the clinical comparison equivalent to its in vitro simulation?

Authors:  B Kleinman; K Frey; R Stevens
Journal:  J Clin Monit Comput       Date:  1998-12       Impact factor: 2.502

Review 3.  Cardiac Output Monitoring by Pulse Contour Analysis, the Technical Basics of Less-Invasive Techniques.

Authors:  Jörn Grensemann
Journal:  Front Med (Lausanne)       Date:  2018-03-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.