Measurement of photon beam backscatter from collimators to the beam monitor chamber using target-current-pulse-counting and telescope techniques.

M K Yu1, R S Sloboda, F Mansour.   

Abstract

Backscattered radiation (BSR) arising from field-defining collimators and entering the beam monitor chamber (BMC) may contribute to observed variations in medical linear accelerator photon beam output with collimator setting. Measuring the magnitude of such contributions for particular accelerators under specified operating conditions is therefore important when attempting to understand and model accelerator head scatter. The present work was conducted to confirm some backscatter measurements for collimating jaws reported previously and to extend these to include other accelerators and a multileaf collimator (MLC). BSR reaching the BMC from the jaws of Clinac 600C, 2100C and 2300CD accelerators and from an MLC on the 2300CD was investigated using both target-current-pulse-counting and telescope methods. Our measurements show that for the Clinac 600C BSR-dependent output variations are negligible. However, for the 2100C and 2300CD BSR-dependent relative output increased in an almost linear fashion, by up to 2.4% for 15 and 18 MV beams, and by up to 1.7% for 6 MV beams, as the field size varied from 5 x 5 cm2 to 40 x 40 cm2. The magnitude of BSR dependent upon collimator location in the head, as expected, thereby contributing to the collimator exchange effect. An earlier study at our centre using the telescope method had reported higher BSR levels. This discrepancy was resolved when corrections for telescope block and room scatter, previously assumed negligible, were made.

Entities:  

Mesh:

Year:  1996        PMID: 8822779     DOI: 10.1088/0031-9155/41/7/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Experimental evaluations of head scatter factor calculation by use of a Gaussian function.

Authors:  Ryoichi Notake; Yukishige Kyuuma; Kouji Sasamori; Makoto Sagawa; Yasushi Annaka; Hidetoshi Saitoh; Hitoshi Shibuya
Journal:  Radiol Phys Technol       Date:  2013-12-10

2.  Prediction of back-scatter radiations to a beam monitor chamber of medical linear accelerators by use of the digitized target-current-pulse analysis method.

Authors:  Yusuke Suzuki; Naoki Hayashi; Hideki Kato; Hiroshi Fukuma; Yasujiro Hirose; Makoto Kawano; Yoshio Nishii; Masaru Nakamura; Takashi Mukouyama
Journal:  Radiol Phys Technol       Date:  2012-10-25

3.  Measurement of back-scattered radiation from micro multileaf collimator into the beam monitor chamber from a dual energy linear accelerator.

Authors:  K R Muralidhar; P Narayana Murthy; N V N M Sresty; Pramod Kumar Dixit; Rajneesh Kumar; A K Raju
Journal:  J Med Phys       Date:  2007-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.