The uneven distribution of the ambient temperature in a model of an infant incubator was demonstrated using the computational fluid mechanical (CFM) simulation of the air flow. A finite volume method of CFM calculation was performed on a three-dimensional (3D) model of an infant incubator including a model baby. The time course of the temperature distribution was computed solving the heat transfer equations simultaneously with the momentum equations. An uneven temperature distribution was observed for a long period (60 s) after the warm inflow was introduced into the incubator chamber. The temperature distribution was complex in 3D space and unsteady even after a long time, suggesting that it may take a considerable time to settle and may continue to be unsteady even if the inflow velocity is steady.Entities:
Mesh:
Year: 1996 PMID: 8803561
Source DB: PubMed Journal: Front Med Biol Eng ISSN: 0921-3775