Comparison of infrared ear thermometer derived and equilibrated rectal temperatures in estimating pulmonary artery temperatures.

L C Rotello1, L Crawford, T E Terndrup.   

Abstract

OBJECTIVES: To investigate the clinical accuracy of infrared ear thermometer derived and equilibrated rectal temperatures in estimating core body temperature. The clinical bias (i.e., mean difference between body sites), and variability (SD of the differences) of simultaneous temperatures were compared with pulmonary artery temperatures. Clinical repeatability (pooled SD of triplicate reading differences) was also examined for three ear infrared thermometers.
DESIGN: Prospective clinical study.
SETTING: A multidisciplinary, adult intensive care unit. PATIENTS: Twenty patients with an existing pulmonary artery catheter were studied in a multidisciplinary, adult intensive care unit.
INTERVENTIONS: A single operator using optimum ear infrared technique and masked to ear and rectal temperatures recorded triplicate measurements with each of three infrared ear thermometers, each over a 4-min period with each infrared thermometer, while an assistant recorded temperatures. Infrared and rectal temperatures were compared with a simultaneous pulmonary artery temperature.
MEASUREMENTS AND MAIN RESULTS: Infrared ear thermometers and rectal thermometers were calibrated daily, and pulmonary artery catheters were calibrated on removal from the patient. Patients were grouped into afebrile and febrile groups, based on initial pulmonary artery temperature. Bias and variability were compared between thermometers using analysis of variance. Clinical bias, but not variability, was significantly different between three ear infrared thermometers (0.16 +/- 0.46 degrees C, 0.07 +/- 0.38 degrees C, and -0.22 +/- 0.47 degrees C). The repeatability was not different between ear infrared thermometers (range 0.13 degrees C to 0.14 degrees C). Rectal temperature had a significantly greater bias (average 0.3 degrees C), but less variability (average 0.2 degrees C). Bias was increased, and variability decreased for both rectal and infrared ear temperatures when pulmonary artery temperature was increased.
CONCLUSIONS: The three infrared ear thermometers studied provided a closer estimate of core body temperature than equilibrated rectal temperature. Clinical bias was greatest in febrile vs. afebrile intensive care unit patients.

Entities:  

Mesh:

Year:  1996        PMID: 8797622     DOI: 10.1097/00003246-199609000-00012

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  9 in total

1.  Is there an association between body temperature and serum lactate levels in hip fracture patients?

Authors:  F Murtuza; A J Farrier; M Venkatesan; R Smith; A Khan; C E Uzoigwe; G Chami
Journal:  Ann R Coll Surg Engl       Date:  2015-08-14       Impact factor: 1.891

2.  Body temperature measurement in paediatrics: Which gadget should we believe?

Authors:  Joan L Robinson
Journal:  Paediatr Child Health       Date:  2004-09       Impact factor: 2.253

3.  Comparison of temporal artery, rectal and esophageal core temperatures in children: Results of a pilot study.

Authors:  Fahad Al-Mukhaizeem; Upton Allen; Luba Komar; Basem Naser; Larry Roy; Derek Stephens; Stanley Read; Christina Kim; Suzanne Schuh
Journal:  Paediatr Child Health       Date:  2004-09       Impact factor: 2.253

4.  Effect of air temperature on the rectal temperature gradient at rest and during exercise.

Authors:  Michael J Buono; Brittany Holloway; Ashley Levine; Cary Rasmussen; Fred W Kolkhorst
Journal:  Int J Physiol Pathophysiol Pharmacol       Date:  2014-03-13

5.  Hypothermia in a surgical intensive care unit.

Authors:  Fernando J Abelha; Maria A Castro; Aida M Neves; Nuno M Landeiro; Cristina C Santos
Journal:  BMC Anesthesiol       Date:  2005-06-06       Impact factor: 2.217

6.  Accuracy of parents in measuring body temperature with a tympanic thermometer.

Authors:  Joan L Robinson; Hsing Jou; Donald W Spady
Journal:  BMC Fam Pract       Date:  2005-01-11       Impact factor: 2.497

7.  Effect evaluation of a heated ambulance mattress-prototype on thermal comfort and patients' temperatures in prehospital emergency care--an intervention study.

Authors:  Jonas Aléx; Stig Karlsson; Ulf Björnstig; Britt-Inger Saveman
Journal:  Int J Circumpolar Health       Date:  2015-09-14       Impact factor: 1.228

8.  Cutaneous infrared thermometry for detecting febrile patients.

Authors:  Pierre Hausfater; Yan Zhao; Stéphanie Defrenne; Pascale Bonnet; Bruno Riou
Journal:  Emerg Infect Dis       Date:  2008-08       Impact factor: 6.883

9.  Emergency department rectal temperatures in over 10 years: A retrospective observational study.

Authors:  Graham A Walker; Daniel Runde; Daniel M Rolston; Dan Wiener; Jarone Lee
Journal:  World J Emerg Med       Date:  2013
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.