Literature DB >> 8797388

Rubber stoppers as sources of contaminants in electrospray analysis of peptides and proteins.

X K Zhang1, R C Dutky, H M Fales.   

Abstract

Using MS/MS, we have identified ions at m/z 399, 421, 609.4, and 819.9 as those of tris (2-butoxyethyl) phosphate, originating in the gray rubber stoppers commonly used as closures in commercial sources of peptides and proteins. The compound is readily extracted by contact with organic solvents such as alcohol or acetic acid and even by water alone. Zinc salts of proteins are also observed, presumably from zinc oxide fillers present in the stoppers.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8797388     DOI: 10.1021/ac960245n

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  Formation of iron complexes from trifluoroacetic acid based liquid chromatography mobile phases as interference ions in liquid chromatography/electrospray ionization mass spectrometric analysis.

Authors:  Anil Shukla; Rui Zhang; Daniel J Orton; Rui Zhao; Therese R W Clauss; Ronald Moore; Richard D Smith
Journal:  Rapid Commun Mass Spectrom       Date:  2011-05-30       Impact factor: 2.419

2.  Identification of dimethyldioctadecylammonium ion (m/z 550.6) and related species (m/z 522.6, 494.6) as a source of contamination in mass spectrometry.

Authors:  M Lisa Manier; D Shannon Cornett; David L Hachey; Richard M Caprioli
Journal:  J Am Soc Mass Spectrom       Date:  2008-02-08       Impact factor: 3.109

3.  A Shaving Proteomic Approach to Unveil Surface Proteins Modulation of Multi-Drug Resistant Pseudomonas aeruginosa Strains Isolated From Cystic Fibrosis Patients.

Authors:  Anna Lisa Montemari; Valeria Marzano; Nour Essa; Stefano Levi Mortera; Martina Rossitto; Simone Gardini; Laura Selan; Gianluca Vrenna; Andrea Onetti Muda; Lorenza Putignani; Ersilia Vita Fiscarelli
Journal:  Front Med (Lausanne)       Date:  2022-03-09
  3 in total