Literature DB >> 8794479

The energy correction factor of LiF thermoluminescent dosemeters in megavoltage electron beams: Monte Carlo simulations and experiments.

P N Mobit1, A E Nahum, P Mayles.   

Abstract

The energy correction factor of LiF thermoluminescent dosemeters (TLDs) calibrated in Co-60 gamma-rays and used for measurements in megavoltage electron beams has been determined experimentally and theoretically using Monte Carlo simulations. The experiments show that the energy correction factor of 1 mm thick TLD-100 has an average for both rods and chips which varies from 1.036 +/- 1.3% (1 SD) for 4 MeV electron beams to 1.021 +/- 1.3% (1 SD) for 20 MeV electron beams for measurement performed at dmax in PMMA (Perspex). The results of the Monte Carlo simulations were within 0.6% of the experimental results and ranged from 1.041 +/- 0.9% (1 SD) for 2 MeV electrons to 1.028 +/- 0.8% (1 SD) for 20 MeV electron beams. There was no significant difference in the energy correction factors of LiF TLDs irradiated in PMMA or water by Monte Carlo simulation and experiments. Differences in the energy correction factors between rods and chips of the same thickness were negligible both in the experiments and in Monte Carlo calculation. When the diameter of the LiF TLD micro-rod was increased from 1 to 5 mm, the simulated energy correction factors increased by as much as 5% over this energy range. The energy correction factors changed by up to 4% for irradiation of TLD at depths other than at dmax for a 5 MeV mono-energetic electron beam.

Entities:  

Mesh:

Year:  1996        PMID: 8794479     DOI: 10.1088/0031-9155/41/6/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Experimental assessment of out-of-field dose components in high energy electron beams used in external beam radiotherapy.

Authors:  Mohamad M Alabdoaburas; Jean-Pierre Mege; Jean Chavaudra; Jérémi Vũ Bezin; Atilla Veres; Florent de Vathaire; Dimitri Lefkopoulos; Ibrahima Diallo
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.