Access flow measurements in hemodialysis patients: in vivo validation of an ultrasound dilution technique.

P J Bosman1, F T Boereboom, C J Bakker, W P Mali, B C Eikelboom, P J Blankestijn, H A Koomans.   

Abstract

In this study, access flow in hemodialysis patients with bridge grafts was measured by a newly developed device. The technique is based on the measurement of changes in the ultrasound characteristics of blood. These changes were initiated by an injection of isotonic saline and measured in the tubes of the extra-corporeal circuit. The access flow rate (mean +/- standard variation) was 880 +/- 440 mL/min (range, 166 to 1740) (N = 46). The mean coefficient of variation was 13.4 +/- 6.8% (median, 13.0%; range 3.5 to 29.4%). Measurements correlated well (r = 0.91, N = 22, P < 0.001) with flow rates determined by magnetic resonance angiography and by a technique based on intra-access flow-pressure curves (r = 0.84, N = 14, P < 0.001). In conclusion, access flow can be measured easily, noninvasively, and reliably by the ultrasound dilution device.

Entities:  

Mesh:

Year:  1996        PMID: 8793808     DOI: 10.1681/ASN.V76966

Source DB:  PubMed          Journal:  J Am Soc Nephrol        ISSN: 1046-6673            Impact factor:   10.121


  2 in total

1.  Early vascular access blood flow as a predictor of long-term vascular access patency in incident hemodialysis patients.

Authors:  Hyung Soo Kim; Jin-Woong Park; Jae Hyun Chang; Jaeseok Yang; Hyun Hee Lee; Wookyung Chung; Yeon Ho Park; Sejoong Kim
Journal:  J Korean Med Sci       Date:  2010-04-22       Impact factor: 2.153

2.  The prototype device for non-invasive diagnosis of arteriovenous fistula condition using machine learning methods.

Authors:  Marcin Grochowina; Lucyna Leniowska; Agnieszka Gala-Błądzińska
Journal:  Sci Rep       Date:  2020-10-02       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.