Thrombosis of mitral valve bioprostheses in patients requiring circulatory assistance.

M T Hagley1, A Lopez-Candales, K J Phillips, B B Daily, N T Kouchoukos.   

Abstract

Two patients with new mitral valve bioprostheses required implantation of Bio-Medicus centrifugal pumps (Bio-Medicus, Minneapolis, MN) for circulatory support and had evidence of prosthetic valve thrombosis 1 and 4 days later. Both patients died of thromboembolic complications despite surgical removal of the thrombus. Thrombosis is a rare early complication of bioprosthetic valves and in these cases was probably related to low transvalvular flow due to the use of circulatory assist devices. We discuss possible strategies for avoiding and managing this catastrophic complication.

Entities:  

Mesh:

Year:  1995        PMID: 8787495     DOI: 10.1016/0003-4975(95)00627-3

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Early acute thrombosis of bioprosthetic mitral valve presenting with cardiogenic shock.

Authors:  Athanasios Tsiouris; Hassan Nemeh; Jamil Borgi
Journal:  Gen Thorac Cardiovasc Surg       Date:  2012-06-20

Review 2.  Successful management of multiple permanent pacemaker complications--infection, 13 year old silent lead perforation and exteriorisation following failed percutaneous extraction, superior vena cava obstruction, tricuspid valve endocarditis, pulmonary embolism and prosthetic tricuspid valve thrombosis.